Agilent Technologies E4406A VSA manual SENSe:ACP:AVERage:TCONtrol EXPonential|REPeat

Models: E4406A VSA

1 406
Download 406 pages 63.49 Kb
Page 287
Image 287
[:SENSe]:ACP:AVERage:TCONtrol EXPonentialREPeat

Language Reference

SENSe Subsystem

Adjacent Channel Power—Averaging Termination Control

[:SENSe]:ACP:AVERage:TCONtrol EXPonentialREPeat

[:SENSe]:ACP:AVERage:TCONtrol?

Select the type of termination control used for averaging. This determines the averaging action after the specified number of data acquisitions (average count) is reached.

EXPonential – Each successive data acquisition after the average count is reached, is exponentially weighted and combined with the existing average.

REPeat – After reaching the average count, the averaging is reset and a new average is started.

Factory Preset

 

and *RST:

REPeat for basic, cdmaOne, cdma2000, W-CDMA

 

(3GPP), W-CDMA (Trial & Arib)

 

EXPonential for NADC, PDC, iDEN

Remarks:

Use INSTrument:SELect to set the mode.

Adjacent Channel Power—Type of Carrier Averaging

[:SENSe]:ACP:AVERage:TYPE MAXimumRMS

[:SENSe]:ACP:AVERage:TYPE?

Selects the type of averaging to be used for the measurement of the carrier.

Factory Preset

 

and *RST:

RMS

Remarks:

You must be in Basic, cdmaOne mode to use this

 

command. Use INSTrument:SELect to set the mode.

History:

Revision A.03.00 or later, in cdmaOne revision A.04.00

Front Panel

 

Access:

Meas Setup, Avg Mode

Adjacent Channel Power—Carrier Channel BW

Basic, cdmaOne, iDEN mode

[:SENSe]:ACP:BANDwidthBWIDth:INTegration <freq>

[:SENSe]:ACP:BANDwidthBWIDth:INTegration?

cdma2000, W-CMDA (3GPP) mode

Chapter 5

287

Page 287
Image 287
Agilent Technologies E4406A VSA manual SENSe:ACP:AVERage:TCONtrol EXPonential|REPeat, SENSe:ACP:AVERage:TCONtrol?