Align the IF Flatness

Language Reference

CALibration Subsystem

The query performs the alignment and returns a zero if the alignment is successful.

Remarks:

A valid service password needs to be entered prior to

 

sending the command.

Front Panel

 

Access:

System, Diagnostics

Align the IF Flatness

:CALibration:FLATness:IF

:CALibration:FLATness:IF?

Finds the flatness shape of the current IF setup (prefilter, mgain, natBW). This information is then used for compensating measurements that use FFT functionality, like the spectrum measurement. The alignment is done frequently in the background. This same alignment is run as part of the CAL:ALL routine.

The query performs the alignment and returns a zero if the alignment is successful.

Front Panel

Access: Select Timebase Freq under Measure, then press Meas Setup, Auto Adjust Now.

Auto Adjust the Internal 10 MHz Frequency Reference

:CALibration:FREQuency:REFerence:AADJust

Auto adjustment of the internal frequency reference (10 MHz timebase). .

Remarks:

You must be in the Service mode to use this command.

 

Use INSTrument:SELect.

 

Requires the current measurement to be timebase

 

frequency. A valid password needs to be entered

 

sometime prior to sending this command. See the

 

timebase frequency measurement for more information.

Front Panel

Select Timebase Freq under Measure, then press Meas

Access:

 

Setup, Auto Adjust Now.