Agilent Technologies E4406A VSA manual Questionable Temperature Enable

Models: E4406A VSA

1 406
Download 406 pages 63.49 Kb
Page 380
Image 380
Questionable Temperature Enable

Language Reference

STATus Subsystem

Questionable Temperature Enable

:STATus:QUEStionable:TEMPerature:ENABle <number>

:STATus:QUEStionable:TEMPerature:ENABle?

This command determines what bits in the Questionable Temperature

Condition Register will set bits in the Questionable Temperature Event

register, which also sets the Temperature Summary bit (bit 4) in the

Questionable Register. The variable <number> is the sum of the

decimal values of the bits you want to enable.

Key Type:

There is no equivalent front panel key.

Factory Preset

 

and *RST:

32767 (all 1’s)

Range:

0 to 32767

Questionable Temperature Event Query :STATus:QUEStionable:TEMPerature[:EVENt]?

This query returns the decimal value of the sum of the bits in the

Questionable Temperature Event register.

NOTE

The register requires that the associated PTR or NTR filters be set

 

before a condition register bit can set a bit in the event register.

 

The data in this register is latched until it is queried. Once queried, the

 

register is cleared

 

Key Type:

There is no equivalent front panel key.

Questionable Temperature Negative Transition

:STATus:QUEStionable:TEMPerature:NTRansition <number>

:STATus:QUEStionable:TEMPerature:NTRansition?

This command determines what bits in the Questionable Temperature Condition register will set the corresponding bit in the Questionable Temperature Event register when the condition register bit has a negative transition (1 to 0). The variable <number> is the sum of the decimal values of the bits that you want to enable.

Key Type:

There is no equivalent front panel key.

Factory Preset

 

and *RST:

0

Range:

0 to 32767

380