Agilent Technologies E4406A VSA manual MHz Amplitude Reference Measurement

Models: E4406A VSA

1 406
Download 406 pages 63.49 Kb
Page 267
Image 267
50 MHz Amplitude Reference Measurement

Language Reference

MEASure Group of Commands

50 MHz Amplitude Reference Measurement

This aligns the internal 50 MHz reference signal to an external reference signal that you supply. You must be in the Service mode to use these commands. Use INSTrument:SELect to set the mode.

The general functionality of CONFigure, FETCh, MEASure, and READ are described at the beginning of this section. See the SENSe:AREFerence commands for more measurement related commands.

:CONFigure:AREFerence

:FETCh:AREFerence[n]?

:READ:AREFerence[n]?

:MEASure:AREFerence[n]?

Remarks:

For auto adjustment of the internal 50 MHz amplitude

 

 

 

reference, use

 

 

 

CALibration:AMPLitude:REFerence:AADJust

 

 

 

command after this measurement has been selected.t

Front Panel

 

Access:

Measure, 50 MHz Amptd

 

 

 

After the measurement is selected, press Restore Meas

 

 

 

Defaults to restore factory defaults.

Measurement Results Available

 

 

n

Results Returned

 

 

not specified or n=1

Returns 7 scalar results:

 

1.

RF input average amplitude

 

2.

50 MHz reference oscillator average amplitude

 

3.

Average amplitude error

 

4.

State (for factory use only)

 

5.

Level (for factory use only)

 

6.

Monitored level (for factory use only)

 

7.

Connector status (for factory use only)

 

 

2

RF input amplitude trace data.

 

 

3

50 MHz oscillator amplitude trace data

 

 

4

Amplitude error strip chart trace data

 

 

 

 

Chapter 5

267

Page 267
Image 267
Agilent Technologies E4406A VSA manual MHz Amplitude Reference Measurement, CONFigure:AREFerence FETCh:AREFerencen?