Agilent Technologies E4406A VSA manual Questionable Calibration Negative Transition

Models: E4406A VSA

1 406
Download 406 pages 63.49 Kb
Page 370
Image 370
Questionable Calibration Negative Transition

Language Reference

STATus Subsystem

NOTE

The register requires that the associated PTR or NTR filters be set

 

before a condition register bit can set a bit in the event register.

 

The data in this register is latched until it is queried. Once queried, the

 

register is cleared.

 

Key Type:

There is no equivalent front panel key.

Questionable Calibration Negative Transition

:STATus:QUEStionable:CALibration:NTRansition <number>

:STATus:QUEStionable:CALibration:NTRansition?

This command determines what bits in the Questionable Calibration Condition register will set the corresponding bit in the Questionable Calibration Event register when the condition register bit has a negative transition (1 to 0). The variable <number> is the sum of the decimal values of the bits that you want to enable.

Key Type:

There is no equivalent front panel key.

Factory Preset

 

and *RST:

0

Range:

0 to 32767

Questionable Calibration Positive Transition

:STATus:QUEStionable:CALibration:PTRansition <number>

:STATus:QUEStionable:CALibration:PTRansition?

This command determines what bits in the Questionable Calibration Condition register will set the corresponding bit in the Questionable Calibration Event register when the condition register bit has a positive transition (0 to 1). The variable <number> is the sum of the decimal values of the bits that you want to enable.

Key Type:

There is no equivalent front panel key.

Factory Preset

 

and *RST:

32767 (all 1’s)

Range:

0 to 32767