Agilent Technologies E4406A VSA Power Statistics CCDF—ChannelBandwidth, SENSe:PSTatistic:COUNts?

Models: E4406A VSA

1 406
Download 406 pages 63.49 Kb
Page 333
Image 333
Power Statistics CCDF—Channel Bandwidth

Language Reference

SENSe Subsystem

Front Panel

Access: Input, Max Total Pwr (at UUT)

Power Statistics CCDF Measurement

Commands for querying the statistical power measurement of the complimentary cumulative distribution function (CCDF) measurement results and for setting to the default values are found in the “MEASure Group of Commands” on page 255. The equivalent front panel keys for the parameters described in the following commands, are found under the Meas Setup key, after the Power Stat CCDF measurement has been selected from the MEASURE key menu.

History:

Added PSTatistic to Basic Mode version A.04.00

Power Statistics CCDF—Channel Bandwidth

[:SENSe]:PSTatistic:BANDwidthBWIDth <freq>

[:SENSe]:PSTatistic:BANDwidthBWIDth?

Set the bandwidth that will be used for acquiring the signal.

Factory Preset

 

and *RST:

5.0 MHz

Range:

10.0 kHz to 6.7 MHz

Resolution:

0.1 kHz

Step:

1.0 kHz

Default Unit:

Hz

Remarks:

You must be in the Basic, cdma2000, W-CDMA (3GPP),

 

or W-CDMA (Trial & Arib) mode to use this command.

 

Use INSTrument:SELect to set the mode.

Power Statistics CCDF—Sample Counts

[:SENSe]:PSTatistic:COUNts <integer>

[:SENSe]:PSTatistic:COUNts?

Set the counts. Measurement stops when the sample counts reach this value.

Factory Preset

 

and *RST:

10,000,000

Range:

1,000 to 2,000,000,000

Unit:

counts

Chapter 5

333