Front Panel Menus | Menu Archive |
Table
Modulation
Measurements
When in the WCDMA measurement mode with the Code Domain Power/Error graph type selected, the following marker selections are available (Refer to the WCDMA Graph Type Dialog on page
Active Marker: To select the active marker, press on the Active Marker field and select a marker from the
To turn the active marker on or off, press the Off/On button. Delta Markers are not available.
Marker n(Code@SF): Set the Code Number@ Spreading Factor by pressing in the Marker n(Code@SF) parameter field (where n is the active marker’s number) and entering a value using the numeric keys. When a new value is entered, the terminator
The marker’s value is displayed in the field below Marker n(Code@SF) and on the graticule display.
All Markers Off: Press the All Markers Off button to deactivate all of the markers.
Uplink Branch: Press the I/Q toggle button to select between the I or Q channel of the WCDMA Uplink mode. This button is only available when in the Uplink mode.
Note: Marker selections in the WiMAX measurement mode are made via the WiMAX application. Refer to the Signature™ Option 41 SignalLab WiMAX Analysis Software User Guide, part number:
Marker to (–>) Menu
The Marker To (–>) menu lets you perform the following actions:
Table
Marker selections are made available depending on the measurement mode and graph type selected.
Next Peak: To send the marker to the next trace peak, press on the Next Peak button.
CF = Mkr: To change the center frequency to the current marker frequency value, press on the CF = Mkr button.
RL = Mkr: To change the reference level to the current marker value, press on the RL = Mkr button.