Preventive Maintenance | General Information |
Replacing the Line Fuses
The line fuses used in the MS278XB are 6.3A, type T fuses. The line fuse values are printed on the rear panel next to the power connector. Always use a new fuse of the type and rating specified by the fuse markings on the rear panel of the instrument. To replace the line fuses, follow the procedure below.
Danger: Before changing the fuse, always remove the power cord from the power outlet. There is the risk of receiving a fatal electric shock if the fuse is replaced with the power cord connected.
Refer to Figure
1.Set the MS278XB to standby mode using the power button and disconnect the power cord from the rear panel power receptacle.
2.Using a small
3.Slide out the fuse holder.
4.Replace the fuse in the fuse holder.
5.Install the fuse holder back into the rear panel fuse block.
6.Close the cover to secure the fuse holder in place. The cover will close with an audible snap.
Figure 1-3. Replacing the Rear Panel Fuse
7.Reconnect the analyzer to the power source and set the MS278XB to Operate using the front panel power button.
Internal Battery
The MS278XB has a Lithium battery installed on the CPU assembly. Battery replacement should be referred to an authorized Anritsu service center.