Measurements | Operation Verification |
5-6 Operation Verification
This section provides procedures that you can use to verify the signal analyzer is operating properly and meets some of its specifications. You may perform the following verifications:
•Reference Oscillator Aging Rate (Optional)
•Frequency Readout Accuracy
•Frequency Span Accuracy
•Swept Resolution Bandwidth
•Single Sideband Phase Noise
•Average Noise Level
•Frequency Response
•Reference Level Switching Uncertainty
•Resolution Bandwidth (RBW) Switching Uncertainty
•Residual Spurious Response
•
Refer to the Required Equipment list before beginning these verification procedures.
Note: Specifications listed in this section are for reference only and should be verified with the Signature Datasheet, part number:
Required Equipment
Table
Table
Instrument | Critical Specification | Recommended Manufacturer/Model |
|
|
|
Frequency Standard | Frequency: 10 MHz | Symmetricom Model |
| Accuracy: 5 ⋅ |
|
Synthesized Signal | Frequency Range: 10 MHz to 8 GHz | Anritsu Model MG3691B with |
Generator | Ultra Low Phase Noise | Options 2A, 3, 4, 16 |
|
|
|
Synthesized Signal | Frequency Range: 10 MHz to 6 GHz | HP 8665B |
Generator | Spurious Performance |
|
|
|
|
Adapter | N(m) to K(f) | Anritsu Model 34NKF50 |
|
|
|
Adapter | Ruggedized K(m) to N(f) | Anritsu Model 34RKNF50 |
|
|
|
Attenuator | K(m) to K(f) | Anritsu Model |
| Frequency Range: DC to 8 GHz |
|
| Attenuation: 3 dB |
|
Attenuator | N(m) to N(f) | Weinschel Model |
| Frequency Range: DC to 8 GHz |
|
| Attenuation: 3 dB |
|
Attenuator | N(m) to N(f) | Weinschel Model 4410 |
| Frequency Range: DC to 8 GHz |
|
| Attenuation: 10 dB |
|
Cable | N(m) to N(m) | Any |
| Frequency Range: DC to 8 GHz |
|
| Impedance: 50 Ω |
|