Basic Measurement Setups

Measurements

Setting the Sweep Time

The Sweep Time sets the duration of a full span sweep. When the sweep time is set to Manual mode, the RBW and VBW settings need to be set to Auto to allow for their automatic determination.

1.Press the BW icon on the main menu bar, or press the Bandwidth key on the instrument front panel, to display the Bandwidth menu below:

2.Select Auto Sweep Time mode or press in the Sweep Time parameter field and use the keypad to enter the desired sweep time, then select the appropriate terminator from the fly-out menu.

5-10

MS278XB OM

Page 166
Image 166
Anritsu Series MS278XB operation manual Setting the Sweep Time