Instrument Overview | Graphical User Interface Overview |
Table
Index Description
Status Indicator: This indicator (asterisk) is displayed when the current instrument setup does not correspond to the one which was applied when a trace is displayed. This happens when:
1The instrument setup is modified when the Sweep mode is set to Single and a new sweep has not yet been triggered.
The instrument setup is modified in the middle of a measurement. The measurement is
2
3Sweep Status: Displays the sweep status of Triggered and Single sweep modes.
Trace Information: Trace information is displayed in the following format (the current active trace has the text underlined):
4 to 8
Trc<n> <Trace Mode>
<Detector Mode>
<Trace Math>
Where:
n:Trace Number (1 to 5)
Trace Mode: Trace Mode (Clear Write, Max Hold, Min Hold, Average, View, or Off)
Detector: Detector Type (Auto, Normal, Maximum, Minimum, Sample, Average, or RMS)
Trace Math: Trace Math indicator (available when Trace Mode: Average is selected)
9 | Sweep Mode: This field is used to display the sweep modes of Swept, FFT (Fast Fourier Transform), or | |
| ||
10 | Start Frequency: Indicates the start frequency of the sweep. | |
|
| |
11 | Frequency/Division: This field indicates the horizontal scaling per division. | |
|
| |
12 | Center Frequency: Indicates the center frequency of the sweep. If a frequency offset is set, OFFSET is | |
also displayed. | ||
| ||
13 | Span Frequency: Indicates the total span of the frequency sweep. | |
|
| |
14 | Stop Frequency: Indicates the stop frequency of the sweep. | |
|
| |
15 | Trace Display: Up to five traces can be displayed using the following colors: Trace | |
Green, Trace | ||
| ||
16 | Marker Display: Markers are annotated with the absolute power level next to the marker location. | |
17 | Marker Display: Markers are annotated with the absolute power level next to the marker location. | |
|
| |
18 | Display Line: A display line may be displayed at any desired level. | |
|
| |
| Active Marker Field: The active marker number, frequency, and amplitude values are displayed in this | |
19 | field. The trace that has received the marker is indicated in brackets. If the Marker is a Delta Marker, the | |
delta symbol (D) precedes the Mkr number and the values displayed are relative to the reference | ||
| ||
| marker. | |
|
| |
20 | Reference Marker Field: The reference marker number, frequency, and amplitude values are displayed | |
in this field. The displayed values are absolute in magnitude. | ||
| ||
|
| |
21 | Bandwidth Setup: The bandwidth setup field displays the current Resolution Bandwidth, Video | |
Bandwidth, and Sweep Time settings. If the settings are manually entered, an asterisk is displayed. | ||
| ||
22 | Amplitude Setup: The amplitude setup field displays the current attenuation level, reference level, and | |
reference level offset. If the attenuation is set manually, an asterisk is displayed. | ||
|