Operation Verification

Measurements

10.Record the maximum Spurious amplitudes in the appropriate column of Table 5-14 based on the frequency offset from the carrier frequency.

Note: Any spurs found should be verified not to originate from the signal source.

Table 5-14.Input-related Spurious Responses Test

Center

Source

Spur Amplitude

 

Spur Amplitude

 

Frequency

Frequency

f<300 kHz from

Specification

f300 kHz from

Specification

(MHz)

(MHz)

Carrier

(dBc)

Carrier

(dBc)

 

 

 

 

 

 

10

10

 

–73

 

–80

 

 

 

 

 

 

12.5

12.5

 

–73

 

–80

 

 

 

 

 

 

16

16

 

–73

 

–80

 

 

 

 

 

 

21.5

21.5

 

–73

 

–80

 

 

 

 

 

 

32

32

 

–73

 

–80

 

 

 

 

 

 

142.5

142.5

 

–73

 

–80

 

 

 

 

 

 

500

500

 

–73

 

–80

 

 

 

 

 

 

650

650

 

–73

 

–80

 

 

 

 

 

 

820

820

 

–73

 

–80

 

 

 

 

 

 

950

950

 

–73

 

–80

 

 

 

 

 

 

999

999

 

–73

 

–80

 

 

 

 

 

 

1020

1020

 

–73

 

–80

 

 

 

 

 

 

2640

2640

 

–73

 

–80

 

 

 

 

 

 

3000

3000

 

–73

 

–80

 

 

 

 

 

 

3500

3500

 

–73

 

–80

 

 

 

 

 

 

4000

4000

 

–73

 

–80

 

 

 

 

 

 

4500

4500

 

–73

 

–80

 

 

 

 

 

 

5000

5000

 

–73

 

–80

 

 

 

 

 

 

5500

5500

 

–73

 

–80

 

 

 

 

 

 

6000

6000

 

–73

 

–80

 

 

 

 

 

 

5-96

MS278XB OM

Page 252