Measurement Examples | Measurements |
Displaying Trace Modes
The following example shows the display of a 1 GHz carrier using the MaxHold, Average, and MinHold trace states:
1.Preset the analyzer using the Preset button.
2.Set the center frequency to 1 GHz and the span to 300 kHz.
3.On the Trace Config menu, set Trace 1 to the MaxHold state.
4.Select Trace 2 as the active trace and set it to the Average state with 5 averages.
5.Select Trace 3 as the active trace and set it to the MinHold state.
The display should look like the one in the figure below:
Figure 5-7. Displaying Trace Modes
MS278XB OM |