Measurements

Measurement Examples

Displaying Detectors

The following example shows an example of a 1 GHz carrier using the Average, RMS, Minimum, and Maximum detectors:

1.Preset the analyzer using the Preset button.

2.Set the center frequency to 1 GHz and the span to 300 kHz.

3.On the Trace Config menu, set the Trace 1 detector to Average.

4.Select Trace 2 as the active trace and set it’s detector to RMS.

5.Select Trace 3 as the active trace and set it’s detector to Minimum.

6.Select Trace 4 as the active trace and set it’s detector to Maximum. The display should look like the one in the figure below:

Figure 5-6.Displaying Detector Modes

MS278XB OM

5-25