Operation Verification

Measurements

9.Turn on Marker and perform Marker Peak search. Read the marker frequency and verify that the value is within specifications using the formula below:

±((marker freq x reference accuracy) + span accuracy + (0.05 x RBW) + (0.5 x last digit))

For technical specifications, refer to the Signature Technical Data Sheet, part number: 11410- 00396, located in Appendix A of the Signature operation manual.

10.Repeat steps 8 through 9 for other RBW/Span combinations listed in Table 5-4.

11.Repeat steps 7 through 10 for 7900 MHz.

Table 5-4.Frequency Readout Accuracy Test

FREQ

RBW

Span

Specification

Marker Readout

(MHz)

(Hz)

(Hz)

(Hz)

Frequency

 

 

 

 

 

100

100

100

 

 

 

 

 

 

 

 

100

1000

 

 

 

 

 

 

 

 

100

10,000

 

 

 

 

 

 

 

 

1000

1,000

 

 

 

 

 

 

 

 

1000

10,000

 

 

 

 

 

 

 

 

1000

100,000

 

 

 

 

 

 

 

 

10,000

10,000

 

 

 

 

 

 

 

 

10,000

100,000

 

 

 

 

 

 

 

 

10,000

1,000,000

 

 

 

 

 

 

 

7900

100

100

 

 

 

 

 

 

 

 

100

1,000

 

 

 

 

 

 

 

 

100

10,000

 

 

 

 

 

 

 

 

1000

1000

 

 

 

 

 

 

 

 

1000

10,000

 

 

 

 

 

 

 

 

1000

100,000

 

 

 

 

 

 

 

 

10,000

10,000

 

 

 

 

 

 

 

 

10,000

100,000

 

 

 

 

 

 

 

 

10,000

1,000,000

 

 

 

 

 

 

 