Measurements | Operation Verification |
5.Press the Sweep button from the Sweep menu and allow averaging to complete. The display should look similar to that below:
Figure 5-54. Displaying the Average Noise Level
6.Record the average displayed noise level in Table
7.Set the Center Frequency of the MS2781B to the next Center Frequency listed in Table
8.Press the Sweep button on the Sweep menu and allow averaging to complete.
9.Record the displayed noise level in Table
10.Repeat steps 7 through 9 for the rest of the center frequencies in Table
MS278XB OM |