Measurements

Operation Verification

5.Press the Sweep button from the Sweep menu and allow averaging to complete. The display should look similar to that below:

Figure 5-54.Displaying the Average Noise Level

6.Record the average displayed noise level in Table 5-9 and verify that it meets the specification.

7.Set the Center Frequency of the MS2781B to the next Center Frequency listed in Table 5-9.

8.Press the Sweep button on the Sweep menu and allow averaging to complete.

9.Record the displayed noise level in Table 5-9 and verify that it meets the specification.

10.Repeat steps 7 through 9 for the rest of the center frequencies in Table 5-9.

MS278XB OM

5-85