Front Panel Menus |
| Parameter Entry and Measurement Setup Dialog Archive | |
|
|
|
|
| ACP Tab (2 of 2) |
| Alt2 Channel: Toggles the Alt2 channel |
|
| measurement off or on. | |
|
|
| |
|
|
| Alt2 Channel Width: Sets the bandwidth of the Alt2 |
|
|
| channel. |
|
|
| Range: 1 Hz to 8 GHz |
|
|
| Default: 5 MHz |
|
|
| Alt2 Channel Spacing: Sets the spacing between |
|
|
| the main and Al2 channel. |
|
|
| Range: 1 Hz to 8 GHz |
|
|
| Default: 15 MHz |
|
|
| RRC Filter: This is used to turn the |
|
|
| filter (RRC) On or Off. |
|
|
| Rolloff Factor: Sets the |
|
|
| measurement (RRC filter must be on). |
|
|
| Range: 0.1 to 1.0 in steps of 0.01 |
|
|
| Default: 0.22. |
|
|
| Bandwidth: Sets the bandwidth for the measurement |
|
|
| (RRC filter must be on). |
|
|
| Range: 1 Hz to 8 GHz |
|
|
| Default: 3.84 MHz. |
|
|
|
|
| Advanced Tab |
|
|
|
|
| /Hz: Toggles the result display between dBm or |
|
|
| dBm/Hz. The measurement type must be set to Abs |
|
|
| before /Hz can be toggled on. |
|
|
| Type: Toggles between an absolute or relative |
|
|
| measurement. |
|
|
| Noise Compensation: When noise compensation is |
|
|
| toggled on, a noise floor measurement is subtracted |
|
|
| from the channel power. |
|
|
| FFT: Toggles between FFT or Swept modes. |
|
|
| Start Time: Sets the start of the measurement |
|
|
| relative to a triggered event (currently not available). |
|
|
| Stop Time: Sets the stop time of the measurement |
|
|
| relative to a triggered event (currently not available). |
|
|
|
|
Figure 4-15. Adjacent Channel Power Setup Dialog