Front Panel Menus

 

Parameter Entry and Measurement Setup Dialog Archive

 

 

 

 

 

ACP Tab (2 of 2)

 

Alt2 Channel: Toggles the Alt2 channel

 

 

measurement off or on.

 

 

 

 

 

 

Alt2 Channel Width: Sets the bandwidth of the Alt2

 

 

 

channel.

 

 

 

Range: 1 Hz to 8 GHz

 

 

 

Default: 5 MHz

 

 

 

Alt2 Channel Spacing: Sets the spacing between

 

 

 

the main and Al2 channel.

 

 

 

Range: 1 Hz to 8 GHz

 

 

 

Default: 15 MHz

 

 

 

RRC Filter: This is used to turn the root-raised-cosine

 

 

 

filter (RRC) On or Off.

 

 

 

Rolloff Factor: Sets the roll-off factor for the

 

 

 

measurement (RRC filter must be on).

 

 

 

Range: 0.1 to 1.0 in steps of 0.01

 

 

 

Default: 0.22.

 

 

 

Bandwidth: Sets the bandwidth for the measurement

 

 

 

(RRC filter must be on).

 

 

 

Range: 1 Hz to 8 GHz

 

 

 

Default: 3.84 MHz.

 

 

 

 

 

Advanced Tab

 

 

 

 

 

/Hz: Toggles the result display between dBm or

 

 

 

dBm/Hz. The measurement type must be set to Abs

 

 

 

before /Hz can be toggled on.

 

 

 

Type: Toggles between an absolute or relative

 

 

 

measurement.

 

 

 

Noise Compensation: When noise compensation is

 

 

 

toggled on, a noise floor measurement is subtracted

 

 

 

from the channel power.

 

 

 

FFT: Toggles between FFT or Swept modes.

 

 

 

Start Time: Sets the start of the measurement

 

 

 

relative to a triggered event (currently not available).

 

 

 

Stop Time: Sets the stop time of the measurement

 

 

 

relative to a triggered event (currently not available).

 

 

 

 

Figure 4-15.Adjacent Channel Power Setup Dialog

MS278XB OM

4-39

Page 137
Image 137
Anritsu Series MS278XB operation manual ACP Tab 2