Operation Verification

Measurements

10.Subtract the new reading from the marker reading at –70 dBm Reference Level setting.

11.Verify that the deviation is within specifications.

12.Repeat steps 8 to 12 for other Reference Level settings listed in Table 5-11.

Table 5-11.Reference Level Switching Uncertainty Test

 

 

Deviation from

 

MS2781B

MS2781B

–70 dBm Reference Level

 

Reference Level Setting

Marker Reading

(dB)

Specifications

 

 

 

 

–70 dBm

 

0

N/A

 

 

 

 

–60 dBm

 

 

0.1 dB

 

 

 

 

–50 dBm

 

 

0.1 dB

 

 

 

 

–40 dBm

 

 

0.1 dB

 

 

 

 

–30 dBm

 

 

0.1 dB

 

 

 

 

–20 dBm

 

 

0.1 dB

 

 

 

 

–10 dBm

 

 

0.1 dB

 

 

 

 

0 dBm

 

 

0.1 dB

 

 

 

 