Measurements | Operation Verification |
Residual Spurious Response
The following test can be used to verify that the MS2781B meets Residual Spurious Response specifications.
Test Setup
Connect the equipment as shown in the figure below:
Figure 5-58. Equipment Setup for Residual Spurious Responses Test
Test Procedure
1.Turn on both the MS2781B Signal Analyzer and warm up for one hour.
2.Press the Preset key on the MS2781B to reset the instrument to factory default state.
3.Set up the MS2781B as follows:
a.Frequency Span: 1 MHz
b.Sweep Type: Swept
c.RBW: 5 kHz
d.VBW: 200 Hz
e.Attenuation: 0 dB
f.Reference Level:
g.Trace Detector: Maximum
4.Set the MS2781B center frequency to the values listed in Table
MS278XB OM |