Operation Verification

Measurements

12.Use the formula below to verify whether the measured result is within specifications:

13.Repeat steps 7 through 12 for the other combinations of Span and RBW settings listed in Table 5-5.

Table 5-5.Frequency Span Accuracy Test

Span

RBW

F1

F2

 

 

Specification

Span Error

(Hz)

(Hz)

(MHz)

(MHz)

F1’

F2’

(Hz)

(F2’ – F1’)/0.8

 

 

 

 

 

 

 

 

10k

100

3999.996

4000.004

 

 

 

 

 

 

 

 

 

 

 

 

100k

1k

3999.96

4000.04

 

 

 

 

 

 

 

 

 

 

 

 

100k

10k

3999.96

4000.04

 

 

 

 

 

 

 

 

 

 

 

 

1M

50k

3999.6

4000.4

 

 

 

 

 

 

 

 

 

 

 

 

10M

100k

3996

4004

 

 

 

 

 

 

 

 

 

 

 

 

33M

100k

3986.8

4013.2

 

 

 

 

 

 

 

 

 

 

 

 

34M

100k

3986.4

4013.6

 

 

 

 

 

 

 

 

 

 

 

 

81M

100k

3967.6

4032.4

 

 

 

 

 

 

 

 

 

 

 

 

500M

20k

3800

4200

 

 

 

 

 

 

 

 

 

 

 

 

500M

1M

3800

4200

 

 

 

 

 

 

 

 

 

 

 

 

1000M

20k

3600

4400

 

 

 

 

 

 

 

 

 

 

 

 

1000M

1M

3600

4400

 

 

 

 

 

 

 

 

 

 

 

 

2000M

50k

3200

4800

 

 

 

 

 

 

 

 

 

 

 

 

2000M

1M

3200

4800

 

 

 

 

 

 

 

 

 

 

 

 

4000M

100k

2400

5600

 

 

 

 

 

 

 

 

 

 

 

 

4000M

1M

2400

5600

 

 

 

 

 

 

 

 

 

 

 

 

8000M

100k

800

7200

 

 

 

 

 

 

 

 

 

 

 

 

8000M

1M

800

7200

 

 

 

 

 

 

 

 

 

 

 

 

5-78

MS278XB OM

Page 234
Image 234
Anritsu Series MS278XB operation manual Specification Span Error MHz F1’ F2’ F2’ F1’/0.8