Operation Verification | Measurements |
12.Use the formula below to verify whether the measured result is within specifications:
13.Repeat steps 7 through 12 for the other combinations of Span and RBW settings listed in Table
Table
Span | RBW | F1 | F2 |
|
| Specification | Span Error |
(Hz) | (Hz) | (MHz) | (MHz) | F1’ | F2’ | (Hz) | (F2’ – F1’)/0.8 |
|
|
|
|
|
|
|
|
10k | 100 | 3999.996 | 4000.004 |
|
|
|
|
|
|
|
|
|
|
|
|
100k | 1k | 3999.96 | 4000.04 |
|
|
|
|
|
|
|
|
|
|
|
|
100k | 10k | 3999.96 | 4000.04 |
|
|
|
|
|
|
|
|
|
|
|
|
1M | 50k | 3999.6 | 4000.4 |
|
|
|
|
|
|
|
|
|
|
|
|
10M | 100k | 3996 | 4004 |
|
|
|
|
|
|
|
|
|
|
|
|
33M | 100k | 3986.8 | 4013.2 |
|
|
|
|
|
|
|
|
|
|
|
|
34M | 100k | 3986.4 | 4013.6 |
|
|
|
|
|
|
|
|
|
|
|
|
81M | 100k | 3967.6 | 4032.4 |
|
|
|
|
|
|
|
|
|
|
|
|
500M | 20k | 3800 | 4200 |
|
|
|
|
|
|
|
|
|
|
|
|
500M | 1M | 3800 | 4200 |
|
|
|
|
|
|
|
|
|
|
|
|
1000M | 20k | 3600 | 4400 |
|
|
|
|
|
|
|
|
|
|
|
|
1000M | 1M | 3600 | 4400 |
|
|
|
|
|
|
|
|
|
|
|
|
2000M | 50k | 3200 | 4800 |
|
|
|
|
|
|
|
|
|
|
|
|
2000M | 1M | 3200 | 4800 |
|
|
|
|
|
|
|
|
|
|
|
|
4000M | 100k | 2400 | 5600 |
|
|
|
|
|
|
|
|
|
|
|
|
4000M | 1M | 2400 | 5600 |
|
|
|
|
|
|
|
|
|
|
|
|
8000M | 100k | 800 | 7200 |
|
|
|
|
|
|
|
|
|
|
|
|
8000M | 1M | 800 | 7200 |
|
|
|
|
|
|
|
|
|
|
|
|
MS278XB OM |