Measurement Performance
Figure
Close shutoff valve
downstream from sensor
D1 calibration |
Fill sensor with D1 fluid |
ProLink Menu > |
Calibration > |
Density cal – Point 1 |
Enter density of D1 fluid |
Do Cal |
Calibration in Progress |
light turns red |
Calibration in Progress |
light turns green |
Close |
D2 calibration |
Fill sensor with D2 fluid |
ProLink Menu > |
Calibration > |
Density cal – Point 2 |
Enter density of D2 fluid |
Do Cal |
Calibration in Progress |
light turns red |
Calibration in Progress |
light turns green |
Close |
Done |
Compensation
Measurement Performance
Troubleshooting
Defaults
Configuration and Use Manual | 105 |