Measurement Performance

Figure 10-8D1 and D2 density calibration – ProLink II

Close shutoff valve

downstream from sensor

D1 calibration

Fill sensor with D1 fluid

ProLink Menu >

Calibration >

Density cal – Point 1

Enter density of D1 fluid

Do Cal

Calibration in Progress

light turns red

Calibration in Progress

light turns green

Close

D2 calibration

Fill sensor with D2 fluid

ProLink Menu >

Calibration >

Density cal – Point 2

Enter density of D2 fluid

Do Cal

Calibration in Progress

light turns red

Calibration in Progress

light turns green

Close

Done

Compensation

Measurement Performance

Troubleshooting

Defaults