Emerson Process Management 2400S manual 8.14.2Configuration procedure, Volume Conc SG, Conc Dens

Models: 2400S

1 188
Download 188 pages 39.36 Kb
Page 88
Image 88
Volume Conc (SG)

Optional Configuration

Table 8-17Derived variables and available process variables continued

 

 

Available process variables

 

 

 

 

 

 

 

Derived variable – ProLink II label

Density at

Standard

Specific

Concentration Net

Net

reference

volume

gravity

mass

volume

and definition

temperature

flow rate

 

flow rate

flow rate

 

 

 

 

 

 

Volume Conc (SG)

Volume concentration derived from specific

 

 

 

 

 

gravity

 

 

 

 

 

The percent volume of solute or of material

 

 

 

 

 

in suspension in the total solution, derived

 

 

 

 

 

from specific gravity

 

 

 

 

 

 

 

 

 

 

 

Conc (Dens)

 

 

Concentration derived from reference

 

 

 

 

 

density

 

 

 

 

 

The mass, volume, weight, or number of

 

 

 

 

 

moles of solute or of material in

 

 

 

 

 

suspension in proportion to the total

 

 

 

 

 

solution, derived from reference density

 

 

 

 

 

 

 

 

 

 

 

Conc (SG)

 

Concentration derived from specific gravity

 

 

 

 

 

The mass, volume, weight, or number of moles of solute or of material in suspension in proportion to the total solution, derived from specific gravity

8.14.2Configuration procedure

Complete configuration instructions for the enhanced density application are provided in the manual entitled Micro Motion Enhanced Density Application: Theory, Configuration, and Use. Because of the complexity of this procedure, Micro Motion recommends using ProLink II for detailed configuration.

If it is necessary to use a DeviceNet tool, refer to the enhanced density manual for application information, and to the manual entitled Micro Motion Model 2400S Transmitters for DeviceNet: Device Profile for complete device profile information.

Basic information on setting up the enhanced density application using a DeviceNet tool is provided in Figure 8-3.

80