Measurement Performance
Figure
Close shutoff valve
downstream from sensor
D1 calibration
Fill sensor with
D1 fluid
D2 calibration
Fill sensor with
D2 fluid
Enter density of
D1 fluid
Start D1
calibration
Monitor status
Object: Calibration object (0x65)
Instance: 1
Attribute ID: 12
Data type: REAL
Service: Set
Object: Calibration object (0x65) Instance: 1
Service: 0x4C
Object: Diagnostics object (0x66)
Instance: 1
Attribute ID: 16
Bit: 14
Service: Get
Enter density of
D2 fluid
Start D2
calibration
Monitor status
Object: Calibration object (0x65)
Instance: 1
Attribute ID: 13
Data type: REAL
Service: Set
Object: Calibration object (0x65) Instance: 1
Service: 0x4D
Object: Diagnostics object (0x66)
Instance: 1
Attribute ID: 16
Bit: 13
Service: Get
No |
| Bit 14 0ff? |
| Yes |
|
|
|
|
No | Bit 13 0ff? |
Yes
Done
106 | Micro Motion® Model 2400S Transmitters for DeviceNet™ |