Bit 14 0ff?

Measurement Performance

Figure 10-9D1 and D2 density calibration – DeviceNet tool

Close shutoff valve

downstream from sensor

D1 calibration

Fill sensor with

D1 fluid

D2 calibration

Fill sensor with

D2 fluid

Enter density of

D1 fluid

Start D1

calibration

Monitor status

Object: Calibration object (0x65)

Instance: 1

Manual background Attribute ID: 12

Data type: REAL

Service: Set

Object: Calibration object (0x65) Manual background Instance: 1

Service: 0x4C

Object: Diagnostics object (0x66)

Instance: 1

Manual background Attribute ID: 16

Bit: 14

Service: Get

Enter density of

D2 fluid

Start D2

calibration

Monitor status

Object: Calibration object (0x65)

Instance: 1

Manual background Attribute ID: 13

Data type: REAL

Service: Set

Object: Calibration object (0x65) Manual background Instance: 1

Service: 0x4D

Object: Diagnostics object (0x66)

Instance: 1

Manual background Attribute ID: 16

Bit: 13

Service: Get

No

 

Bit 14 0ff?

 

Yes

 

 

 

 

No

Bit 13 0ff?

Yes

Done