Emerson Process Management 2400S manual •Using ProLink II, see Figure B-3

Models: 2400S

1 188
Download 188 pages 39.36 Kb
Page 77
Image 77
Manual background

Optional Configuration

In addition, if you are using ProLink II or a DeviceNet tool, you can control the intensity of the backlight. You can specify any value between 0 and 63; the higher the value, the brighter the backlight. To control the intensity of the backlight:

Using ProLink II, see Figure B-3.

Using a DeviceNet tool, see Table C-9.

8.9.5Configuring the display variables and display precision

The display can scroll through up to 15 process variables in any order. You can configure the process variables to be displayed and the order in which they should appear. Additionally, you can configure display precision for each display variable. Display precision controls the number of digits to the right of the decimal place. Precision can be set to any value from 0 to 5.

To configure display variables or display precision using ProLink II, see Figure B-3.

To configure display variables using a DeviceNet tool, see Table C-9.

To configure display precision using a DeviceNet tool, refer to Table C-9and:

a.Set the process variable index (Attribute 29) to the process variable to be configured.

b.Set the precision (Attribute 30) for that process variable.

Note: This functionality is not available via the display menus.

Table 8-10shows an example of a display variable configuration. Notice that you can repeat variables, and you can also specify None for any display variable except Display Var 1. For information on how the display variables will appear on the display, see Appendix D.

Table 8-10Example of a display variable configuration

Display variable

Process variable

 

 

Display variable 1(1)

Mass flow

 

 

Display variable 2

Mass totalizer

 

 

Display variable 3

Volume flow

 

 

Display variable 4

Volume totalizer

 

 

Display variable 5

Density

 

 

Display variable 6

Temperature

 

 

Display variable 7

External temperature

 

 

Display variable 8

External pressure

 

 

Display variable 9

Mass flow

 

 

Display variable 10

None

 

 

Display variable 11

None

 

 

Display variable 12

None

 

 

Display variable 13

None

 

 

Display variable 14

None

 

 

Display variable 15

None

 

 

(1) Display variable 1 cannot be set to None.

Using a DeviceNet Tool

Required Configuration

Using the Transmitter

Optional Configuration

Configuration and Use Manual

69

Page 77
Image 77