Troubleshooting
Table
Alarm | ProLink II |
|
|
code | message | Cause | Suggested remedy |
|
|
|
|
A030 | Incorrect Board | The loaded software is not | • Contact Micro Motion. See Section 11.3. |
| Type | compatible with the |
|
|
| programmed board type |
|
|
|
|
|
A031 | Low Power | The transmitter is not | • Check power supply to transmitter. See Section 11.4. |
|
| receiving enough power |
|
|
|
|
|
A032 | Meter | Meter verification in | • Allow the procedure to complete. |
| Verification/Outputs | progress, with outputs set | • If desired, abort the procedure and restart with outputs set |
| In Fault | to fault | to last measured value. |
|
|
|
|
A033 | Sensor OK, Tubes | No signal from LPO or | • Verify process. Check for air in the flow tubes, tubes not |
| Stopped by | RPO, suggesting that | filled, foreign material in tubes, or coating in tubes (see |
| Process | sensor tubes are not | Section 11.15). |
|
| vibrating |
|
|
|
|
|
A102 | Drive Overrange/ | The drive power | • Excessive drive gain. See Section 11.19.3. |
| Partially Full Tube | (current/voltage) is at its | • Check the sensor circuitry. See Section 11.20. |
|
| maximum | • If this is the only active alarm, it can be ignored. If desired, |
|
|
| reconfigure the alarm severity to Ignore (see Section 8.8). |
|
|
|
|
A104 | Calibration in | A calibration procedure is | • Allow the flowmeter to complete calibration. |
| Progress | in progress | • For zero calibration procedures, you may abort the |
|
|
| calibration, set the zero time parameter to a lower value, |
|
|
| and restart the calibration. |
|
|
|
|
A105 | Slug Flow | The density has exceeded | • See Section 11.14. |
|
| the |
|
|
| (density) limits |
|
|
|
|
|
A107 | Power Reset | The transmitter has been | • No action required. |
| Occurred | restarted | • If desired, reconfigure the alarm severity to Ignore (see |
|
|
| Section 8.8). |
|
|
|
|
A116 | API: Temperature | The process temperature | • Verify process. |
| Outside Standard | is outside | • Verify API reference table and temperature configuration. |
| Range | extrapolation limits | See Section 8.13. |
|
|
|
|
A117 | API: Density | The process density is | • Verify process. |
| Outside Standard | outside | • Verify API reference table and density configuration. See |
| Range | extrapolation limits | Section 8.13. |
|
|
|
|
A120 | ED: Unable to Fit | The configured values for | • Verify enhanced density configuration. See Section 8.14. |
| Curve Data | density curves do not |
|
|
| meet accuracy |
|
|
| requirements |
|
|
|
|
|
A121 | ED: Extrapolation | Enhanced density | • Verify process temperature. |
| Alarm | calculations are outside | • Verify process density. |
|
| the configured data range | • Verify enhanced density configuration. See Section 8.14. |
|
|
|
|
A131 | Meter | Meter verification in | • Allow the procedure to complete. |
| Verification/Outputs | progress, with outputs set | • If desired, abort the procedure and restart with outputs set |
| at Last Value | to last measured value | to fault. |
|
|
|
|
A132 | Simulation Mode | Simulation mode is | • Disable simulation mode. See Section 11.10. |
| Active | enabled |
|
|
|
|
|
A133 | PIC UI EEPROM | EEPROM data on the user | • Contact Micro Motion. See Section 11.3. |
| Error | interface module is corrupt |
|
|
|
|
|
118 | Micro Motion® Model 2400S Transmitters for DeviceNet™ |