Troubleshooting

Table 11-2Status alarms and remedies continued

Alarm

ProLink II

 

 

code

message

Cause

Suggested remedy

 

 

 

 

A030

Incorrect Board

The loaded software is not

• Contact Micro Motion. See Section 11.3.

 

Type

compatible with the

 

 

 

programmed board type

 

 

 

 

 

A031

Low Power

The transmitter is not

• Check power supply to transmitter. See Section 11.4.

 

 

receiving enough power

 

 

 

 

 

A032

Meter

Meter verification in

• Allow the procedure to complete.

 

Verification/Outputs

progress, with outputs set

• If desired, abort the procedure and restart with outputs set

 

In Fault

to fault

to last measured value.

 

 

 

 

A033

Sensor OK, Tubes

No signal from LPO or

• Verify process. Check for air in the flow tubes, tubes not

 

Stopped by

RPO, suggesting that

filled, foreign material in tubes, or coating in tubes (see

 

Process

sensor tubes are not

Section 11.15).

 

 

vibrating

 

 

 

 

 

A102

Drive Overrange/

The drive power

• Excessive drive gain. See Section 11.19.3.

 

Partially Full Tube

(current/voltage) is at its

• Check the sensor circuitry. See Section 11.20.

 

 

maximum

• If this is the only active alarm, it can be ignored. If desired,

 

 

 

reconfigure the alarm severity to Ignore (see Section 8.8).

 

 

 

 

A104

Calibration in

A calibration procedure is

• Allow the flowmeter to complete calibration.

 

Progress

in progress

• For zero calibration procedures, you may abort the

 

 

 

calibration, set the zero time parameter to a lower value,

 

 

 

and restart the calibration.

 

 

 

 

A105

Slug Flow

The density has exceeded

• See Section 11.14.

 

 

the user-defined slug

 

 

 

(density) limits

 

 

 

 

 

A107

Power Reset

The transmitter has been

• No action required.

 

Occurred

restarted

• If desired, reconfigure the alarm severity to Ignore (see

 

 

 

Section 8.8).

 

 

 

 

A116

API: Temperature

The process temperature

• Verify process.

 

Outside Standard

is outside API-defined

• Verify API reference table and temperature configuration.

 

Range

extrapolation limits

See Section 8.13.

 

 

 

 

A117

API: Density

The process density is

• Verify process.

 

Outside Standard

outside API-defined

• Verify API reference table and density configuration. See

 

Range

extrapolation limits

Section 8.13.

 

 

 

 

A120

ED: Unable to Fit

The configured values for

• Verify enhanced density configuration. See Section 8.14.

 

Curve Data

density curves do not

 

 

 

meet accuracy

 

 

 

requirements

 

 

 

 

 

A121

ED: Extrapolation

Enhanced density

• Verify process temperature.

 

Alarm

calculations are outside

• Verify process density.

 

 

the configured data range

• Verify enhanced density configuration. See Section 8.14.

 

 

 

 

A131

Meter

Meter verification in

• Allow the procedure to complete.

 

Verification/Outputs

progress, with outputs set

• If desired, abort the procedure and restart with outputs set

 

at Last Value

to last measured value

to fault.

 

 

 

 

A132

Simulation Mode

Simulation mode is

• Disable simulation mode. See Section 11.10.

 

Active

enabled

 

 

 

 

 

A133

PIC UI EEPROM

EEPROM data on the user

• Contact Micro Motion. See Section 11.3.

 

Error

interface module is corrupt

 

 

 

 

 