Troubleshooting
11.12Status alarms
Status alarm codes are reported on the LCD panel (for transmitters that have a display), and status alarms can be viewed with ProLink II or a DeviceNet tool (see Section 7.6). All possible status alarms are listed in Table
You may find it useful to acknowledge all alarms before beginning the troubleshooting procedures. This will remove inactive alarms from the list and allow you to focus on active alarms.
Table 11-2 Status alarms and remedies
Alarm | ProLink II |
|
|
code | message | Cause | Suggested remedy |
|
|
|
|
A001 | (E)EPROM | An uncorrectable | • Cycle power to the flowmeter. |
| Checksum Error | checksum mismatch has | • The flowmeter might need service. Contact Micro Motion. |
| (CP) | been detected | See Section 11.3. |
|
|
|
|
A002 | RAM Error (CP) | ROM checksum error or a | • Cycle power to the flowmeter. |
|
| RAM location cannot be | • The flowmeter might need service. Contact Micro Motion. |
|
| written to | See Section 11.3. |
|
|
|
|
A003 | Sensor Failure | Continuity failure of drive | • Check for slug flow. See Section 11.14. |
|
| circuit, LPO, or RPO, or | • Check the test points. See Section 11.19. |
|
| • Check the sensor circuitry. See Section 11.20. | |
|
| driving | • Check sensor tubes for plugging. |
|
|
| • If the problem persists, contact Micro Motion. See |
|
|
| Section 11.3. |
|
|
|
|
A004 | Temperature | Combination of A016 and | • Check the sensor RTD circuitry. See Section 11.20. |
| Sensor Failure | A017 | • Verify that process temperature is within range of sensor |
|
|
| and transmitter. |
• If the problem persists, contact Micro Motion. See Section 11.3.
A005 | Input Overrange | The measured flow has |
|
| exceeded the maximum |
flow rate of the sensor (ΔT > 200 μs)
•If other alarms are present (typically, A003, A006, A008, A102, or A105), resolve those alarm conditions first. If the A005 alarm persists, continue with the suggestions here.
•Verify process and check for slug flow. See Section 11.14.
•Check the test points. See Section 11.19.
•Check the sensor circuitry. See Section 11.20.
•Check the sensor tubes for erosion. See Section 11.15.
•If the problem persists, contact Micro Motion. See Section 11.3.
A006 | Not Configured | Combination of A020 and | • Check the characterization. Specifically, verify the FCF |
|
| A021 | and K1 values. See Section 6.2. |
|
|
| • If the problem persists, contact Micro Motion. See |
|
|
| Section 11.3. |
|
|
|
|
A008 | Density Overrange | The measured density has | • If other alarms are present (typically, A003, A006, A102, or |
|
| exceeded | A105), resolve those alarm conditions first. If the A008 |
|
|
| alarm persists, continue with the suggestions here. |
• Verify process. Check for air in the flow tubes, tubes not filled, foreign material in tubes, or coating in tubes (see Section 11.15).
• Check for slug flow. See Section 11.14.
• Check the sensor circuitry. See Section 11.20.
• Verify calibration factors in transmitter configuration. See Section 6.2.
• Check the test points. See Section 11.19.
• If the problem persists, contact Micro Motion. See Section 11.3.