Measurement Performance
Figure
Step 1
Set output state (optional)
Step 2
Set uncertainty limit (optional)
See Table
Step 3 |
|
|
|
| |
|
| Manual abort (optional) |
| ||
Start/abort procedure |
|
|
| ||
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
Step 4
Check current algorithm state
No (<16)
Step 9
Check abort code
|
|
|
|
| Step 5 |
|
Running? |
| Yes (>0) |
|
| ||
|
|
| Read percent complete |
| ||
|
|
|
|
|
| |
|
|
|
|
|
|
|
No (=0)
Step 6
Check algorithm abort state
Able to |
|
|
|
|
|
|
|
|
|
|
|
|
| Yes (=16) |
|
|
|
|
|
|
|
|
| ||
complete? |
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
| Step 7 |
|
|
|
|
| |||
|
| Check inlet stiffness |
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Within limits? |
|
|
|
|
| CAUTION | ||||
|
|
|
| No (>0) |
|
| ||||||
|
|
|
|
| ||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
| ||
|
| Yes | (=0) |
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |||
|
|
|
| Step 8 |
|
|
|
|
| |||
|
| Check outlet stiffness |
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
Within limits? |
|
|
|
| CAUTION |
| No (>0) |
| |||
|
| ||||
|
|
|
|
|
|
Yes (=0)
PASS
94 | Micro Motion® Model 2400S Transmitters for DeviceNet™ |