Pressure Compensation and Temperature Compensation
There are two ways to configure external temperature compensation:
•If the operating temperature is a known static value, you can configure that value in the transmitter.
•If the operating temperature varies significantly, you must write a temperature value to the transmitter at appropriate intervals, using an appropriate output assembly. See Section 9.4.
Note: Ensure that your temperature value is accurate, or that your temperature measurement device is accurate and reliable.
To enable and configure external temperature compensation:
•With ProLink II, see Figure
•With a DeviceNet tool, see Figure
Figure
Enable |
View Menu > |
Preferences |
Enable Use External |
Temperature |
Apply |
(1)Temperature measurement unit must be configured to match temperature unit used by external device. See Section 6.3.
(2)See Section 9.4.
Configure |
|
Use static |
|
temp value? |
|
No | Yes |
ProLink > | Enter External |
Configuration > | Temperature |
Temperature |
|
| Apply |
Enter Temperature units(1) |
|
Apply |
|
Set up output |
|
assembly(2) |
|
Done |
|
86 | Micro Motion® Model 2400S Transmitters for DeviceNet™ |