Emerson Process Management 2400S manual Value

Models: 2400S

1 188
Download 188 pages 39.36 Kb
Page 153
Image 153
Manual background

Device Profile

Table C-2Analog Input Point Object (0x0A) – Instance 2 (liquid volume flow)

Attrib

 

 

 

 

 

 

ID

Name

Data type

Service

Mem

Description

Comments

 

 

 

 

 

 

 

3

Value

REAL

Get

V

Current value of liquid

Based on Attribute 8

 

 

 

 

 

volume flow process

 

 

 

 

 

 

variable

 

 

 

 

 

 

 

 

4

Status

BOOL

Get

V

Point status

• 0 = Good

 

 

 

 

 

 

• 1 = Alarm state

 

 

 

 

 

 

 

8

Value data type

USINT

Get

V

Data type used to report

1 (REAL)

 

 

 

 

 

volume flow process

 

 

 

 

 

 

variable

 

 

 

 

 

 

 

 

100

Process total

REAL

Get

V

Current value of liquid

 

 

 

 

Reset(1)

 

volume total

 

 

 

 

 

 

 

 

101

Inventory total

REAL

Get

V

Current value of liquid

 

 

 

 

Reset(2)

 

volume inventory

 

 

 

 

 

 

 

 

102

Value

UINT

Set

NV

Liquid volume flow

See Table 6-3for unit

 

engineering

 

 

 

measurement unit

codes.

 

units

 

 

 

 

 

 

 

 

 

 

 

 

103

Total

UINT

Get

V

Liquid volume total and

Transmitter

 

engineering

 

 

 

inventory units

automatically

 

units

 

 

 

 

determines this based

 

 

 

 

 

 

on Attribute 102.

 

 

 

 

 

 

See Table C-13for unit

 

 

 

 

 

 

codes.

 

 

 

 

 

 

 

105

Cutoff

REAL

Set

NV

Value below which liquid

 

 

 

 

 

 

volume flow will be

 

 

 

 

 

 

reported as 0

 

 

 

 

 

 

 

 

106

Meter factor

REAL

Set

NV

A multiplier to the

 

 

 

 

 

 

calculated liquid volume

 

 

 

 

 

 

flow

 

 

 

 

 

 

 

 

108

Reset volume

USINT

Set

V

Resets the volume total

• 1 = Reset

 

total

 

 

 

 

 

 

 

 

 

 

 

 

109

Reset volume

USINT

Set

V

Resets the volume

• 1 = Reset

 

inventory

 

 

 

inventory

 

(1)Service code 0x4B.

(2)Service code 0x4C.

Table C-3Analog Input Point Object (0x0A) – Instance 3 (density)

Attrib

 

 

 

 

 

 

ID

Name

Data type

Service

Mem

Description

Comments

 

 

 

 

 

 

 

3

Value

REAL

Get

V

Current value of density

Based on Attribute 8

 

 

 

 

 

process variable

 

 

 

 

 

 

 

 

4

Status

BOOL

Get

V

Point status

• 0 = Good

 

 

 

 

 

 

• 1 = Alarm state

 

 

 

 

 

 

 

8

Value data type

USINT

Get

V

Data type used to report

1 (REAL)

 

 

 

 

 

density process variable

 

 

 

 

 

 

 

 

102

Value

UINT

Set

NV

Density measurement unit

See Table 6-5for unit

 

engineering

 

 

 

 

codes.

 

units

 

 

 

 

 

 

 

 

 

 

 

 

Menus

Device Profile

Display Codes

Index

Configuration and Use Manual

145

Page 153
Image 153
Emerson Process Management 2400S manual Value