Optional Configuration

Table 8-3Valid damping values

Process variable

Valid damping values

 

 

Flow (mass and volume)

0, 0.04, 0.08, 0.16, ... 40.96

 

 

Density

0, 0.04, 0.08, 0.16, ... 40.96

 

 

Temperature

0, 0.6, 1.2, 2.4, 4.8, ... 76.8

 

 

To configure damping values:

Using ProLink II, see Figure B-2.

Using a DeviceNet tool, see Tables C-1,C-3,and C-4.

Note: This functionality is not available via the display menus.

8.4.1Damping and volume measurement

When configuring damping values, note the following:

Liquid volume flow is derived from mass and density measurements; therefore, any damping applied to mass flow and density will affect liquid volume measurement.

Gas standard volume flow is derived from mass flow measurement, but not from density measurement. Therefore, only damping applied to mass flow will affect gas standard volume measurement.

Be sure to set damping values accordingly.

8.5Configuring the flow direction parameter

The flow direction parameter controls how the transmitter reports flow rate and how flow is added to or subtracted from the totalizers, under conditions of forward flow, reverse flow, or zero flow.

Forward (positive) flow moves in the direction of the arrow on the sensor.

Reverse (negative) flow moves in the direction opposite of the arrow on the sensor.

Options for flow direction include:

Forward only

Reverse only

Absolute value

Bidirectional

Negate/Forward only

Negate/Bidirectional

For the effect of flow direction on flow totals and flow values, see Table 8-4.

60

Micro Motion® Model 2400S Transmitters for DeviceNet

Page 68