Optional Configuration
5.Click Next.
6.Verify the reference temperature and reference pressure. If these are not appropriate for your application, click the Change Reference Conditions button and enter new values for reference temperature and reference pressure.
7.Click Next. The calculated standard density value is displayed.
•If the value is correct, click Finish. The value will be written to transmitter configuration.
•If the value is not correct, click Back and modify input values as required.
Note: The Gas Wizard displays density, temperature, and pressure in the configured units. If required, you can configure the transmitter to use different units. See Section 6.3.
8.2.2Using a DeviceNet tool
The Gas Standard Volume Object is used to configure volume flow measurement for gas. See the flowchart in Figure
Figure
Enable gas standard
volume flow
measurement
Set unit
Class: Gas Standard Volume Object (0x64)
Instance: 1
Attribute ID: 7
Data type: BOOL Value:
•0 = disabled (and liquid volume flow enabled)
•1 = enabled (and liquid volume flow disabled) Service: Set
Class: Gas Standard Volume Object (0x64)
Instance: 1
Attribute ID: 5
Data type: UINT
Value: See Table
Service: Set
Set cutoff(1)
Set reference density
of gas(2)
Class: Gas Standard Volume Object (0x64)
Instance: 1
Attribute ID: 8
Data type: REAL
Service: Set
Class: Gas Standard Volume Object (0x64)
Instance: 1
Attribute ID: 4
Data type: REAL
Service: Set
(1)See Section 8.3.
(2)The Gas Wizard is provided only with ProLink II. If you are not using ProLink II, you must supply the required reference density.
8.3Configuring cutoffs
Cutoffs are
See Table
58 | Micro Motion® Model 2400S Transmitters for DeviceNet™ |