Using the Transmitter

Table 7-3Module LED states, definitions, and recommendations

Module LED state

Definition

Recommendations

 

 

 

Off

No power

Check the connection to the DeviceNet network.

 

 

 

Solid green

No processor faults

No action required.

 

 

 

Flashing green

Needs DeviceNet configuration;

Indicates an A006 alarm. Characterization

 

may be in Standby state

parameters are missing. See Section 6.2.

 

 

 

Solid red

Non-recoverable fault

Power cycle the transmitter. If condition does not

 

 

clear, call Micro Motion customer service.

 

 

 

Flashing red

Recoverable fault

Check for any status alarms.

 

 

 

Flashing red/green

Device in self-test

Wait until self-test is complete.

 

 

Check the Identity Object (0x01) for device states.

 

 

 

7.4.2Using the network LED

The behavior of the network LED is standard, and is defined by the DeviceNet protocol. Table 7-4lists the different states of the network LED and defines each state.

Table 7-4Network LED states, definitions, and recommendations

Network LED state

Definition

Recommendations

 

 

 

Off

Device not online

The device is not connected to the network.(1) Check

 

 

the wiring if this LED is lit.

 

 

 

Solid green

Device online and connected

No action required.

 

 

 

Flashing green

Device online but not connected

The device is connected to the network, but has not

 

 

been allocated by a host. No action required.

 

 

 

Solid red

Critical link failure

The most common cause is duplicate MAC IDs

 

 

(node addresses) on the network. Check for

 

 

duplicate MAC IDs.

 

 

Other causes include incorrect baud rate setting or

 

 

other network failure.

 

 

 

Flashing red

Connection timeout

Power cycle the device, or release and re-allocate

 

 

the device from the DeviceNet master.

 

 

If desired, increase the timeout value (Expected

 

 

Packet Rate) in the DeviceNet Object (0x03).

 

 

 

Flashing red/green

Communication faulted state

Not implemented in the Model 2400S DN transmitter.

 

 

 

(1)If the transmitter is the only device on the network, and there is no host on the network, this is the expected LED state, and no action is required.

7.5Viewing transmitter status

You can view transmitter status using the status LED, ProLink II, or a DeviceNet tool. Depending on the method chosen, different information is displayed.

7.5.1Using the status LED

The status LED shows transmitter status as described in Table 7-5.Note that the status LED does not report event status or alarm status for alarms with severity level set to Ignore (see Section 8.8).

42