Emerson Process Management 2400S 3Configuring the enhanced density application DeviceNet tool

Models: 2400S

1 188
Download 188 pages 39.36 Kb
Page 89
Image 89

Optional Configuration

Figure 8-3Configuring the enhanced density application – DeviceNet tool

 

 

 

 

Class: Enhanced Density Object (0x64)

 

 

 

 

Instance: 1

 

 

 

 

Attribute ID: 39

 

 

 

 

Enable the enhanced density application

 

 

Data type: BOOL

 

 

Value:

 

 

 

 

 

 

 

 

· 0: Disabled

 

 

 

 

 

 

 

 

· 1: Enabled

 

 

 

 

Service: Set

 

 

 

 

 

 

 

 

 

 

Set transmitter density measurement unit to

 

 

Class: Analog Input Point Object (0x0A)

match curve unit

 

 

Instance: 3

· For standard curves, see Table 8-16

 

 

Attribute ID: 102

 

· For custom curves, see the information provided

 

 

Value: See Table 6-5

with the curve

 

 

Service: Set

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Set transmitter temperature measurement unit to

 

 

Class: Analog Input Point Object (0x0A)

match curve unit

 

 

Instance: 4

· For standard curves, see Table 8-16

 

 

Attribute ID: 102

 

· For custom curves, see the information provided

 

 

Value: See Table 6-6

with the curve

 

 

Service: Set

 

 

 

 

 

Using a DeviceNet Tool

Required Configuration

Set derived variable

·For standard curves, use Mass Conc (Dens)

·For custom curves, see the information provided with the curve

Specify the active curve

Class: Enhanced Density Object (0x64)

Instance: 1

Attribute ID: 15

Data type: USINT

Value: See Table D-17

Service: Set

Class: Enhanced Density Object (0x64)

Instance: 1

Attribute ID: 16

Data type: USINT

Value: 0 – 5

Service: Set

Using the Transmitter

Optional Configuration

Configuration and Use Manual

81

Page 89
Image 89
Emerson Process Management 2400S manual 3Configuring the enhanced density application DeviceNet tool