5.5Default assemblies

Using a DeviceNet Tool

5.5Default assemblies

The default assemblies used by the Model 2400S DN transmitter are listed and described in Table 5-1.To change the default assemblies, see the flowchart in Figure 5-1.

Table 5-1Default DeviceNet assemblies

Connection type

Assembly type

Instance ID

Description

Size (bytes)

Data type

 

 

 

 

 

 

Polled

Input

6

Status

21

BOOL

 

 

 

Mass flow

 

REAL

 

 

 

Mass total

 

REAL

 

 

 

Mass inventory

 

REAL

 

 

 

Temperature

 

REAL

 

 

 

Density

 

REAL

 

 

 

 

 

 

 

Output

54

Reset all totalizer

1

BOOL

 

 

 

values

 

 

 

 

 

 

 

 

Cyclic

Input

6

Status

21

BOOL

 

 

 

Mass flow

 

REAL

 

 

 

Mass total

 

REAL

 

 

 

Mass inventory

 

REAL

 

 

 

Temperature

 

REAL

 

 

 

Density

 

REAL

 

 

 

 

 

 

Figure 5-1Changing the default DeviceNet assemblies

Using a DeviceNet Tool

Required Configuration

Polled connection:

Input assembly

Polled connection: Output assembly

Cyclic connection:

Input assembly

Class: Connection Object (0x95)

Instance: 1

Attribute ID: 100

Data type: UINT

Value: See Table 7-2

Service: Set

Class: Connection Object (0x95)

Instance: 1

Attribute ID: 101

Data type: UINT

Value: See Tables 7-8 and 9-1

Service: Set

Class: Connection Object (0x95)

Instance: 1

Attribute ID: 102

Data type: UINT

Value: See Table 7-2

Service: Set

Using the Transmitter

Optional Configuration

Configuration and Use Manual

23