Using the Transmitter

To view information about a single alarm:

1.Execute a Set for Attribute 18, specifying the code for the alarm you want to check.

2.Execute a Get for Attribute 42, and interpret the data using the following codes:

0x00 = Acknowledged and cleared

0x01 = Active and acknowledged

0x10 = Not acknowledged, but cleared

0x11 = Not acknowledged, and active

3.Other information about the indexed alarm is available in the following attributes:

Attribute 43: Number of times this alarm has become active

Attribute 44: The time this alarm was last posted

Attribute 45: The time this alarm was last cleared

To acknowledge an alarm:

1.Execute a Set for Attribute 18, specifying the code for the alarm you want to acknowledge.

2.Execute a Set for Attribute 42, specifying a value of 0x00.

To retrieve information from alarm history:

1.Execute a Set for Attribute 46, specifying the number of the alarm record you want to check. Valid values are 0–49.

Note: The alarm history is a circular buffer, and older records are overwritten by newer records. To determine whether a record is newer or older than another record, you must compare their timestamps.

2.Execute Gets for the following attributes:

Attribute 47: The alarm type

Attribute 49: The time that this alarm changed status

Attribute 48: The type of status change:

-1 = Alarm posted

-2 = Alarm cleared

7.7Using the totalizers and inventories

The totalizers keep track of the total amount of mass or volume measured by the transmitter over a period of time.

The inventories track the same values as the totalizers. Whenever totalizers are started or stopped, all inventories (including the API volume inventory and enhanced density inventories) are started or stopped automatically. However, when totalizers are reset, inventories are not reset automatically – you must reset inventories separately. This allows you to use the inventories to keep running totals across multiple totalizer resets.

You can view all totalizer and inventory values using any of the commmunication tools: the display, ProLink II, or a DeviceNet tool. Specific starting, stopping, and resetting functionality depends on the tool you are using.

Using a DeviceNet Tool

Required Configuration

Using the Transmitter

Optional Configuration

Configuration and Use Manual