Optional Configuration
3.Assign a process variable to the event (Attribute 10).
4.Specify the event’s setpoint(s) – the value(s) at which the event will occur or switch state (ON to OFF, or vice versa).
•If Event Type is High or Low, only Setpoint A is used (Attribute 8)
•If Event Type is In Range or Out of Range, both Setpoint A and Setpoint B (Attributes 9 and 10) are required.
5.Assign the event to an action or actions, if desired. Possible actions are listed in Table
•Using ProLink II, open the Discrete Input panel in the Configuration window, identify the action to be performed, then specify the event using the dropdown list. See Figure
Note: For consistency with other Micro Motion products, the Discrete Input panel is used here even though the Model 2400S DN transmitter does not provide a discrete input.
•Using the display, see Figure
•Using a DeviceNet tool, refer to Table
Table 8-5 Event types
| DeviceNet |
|
Type | code | Description |
|
|
|
High (> A) | 0 | Default. Discrete event will occur if the assigned variable is greater than the |
|
| setpoint (A).(1) |
|
|
|
Low (< A) | 1 | Discrete event will occur if the assigned variable is less than the setpoint (A).(1) |
|
|
|
In Range | 2 | Discrete event will occur if the assigned variable is greater than or equal to the low |
|
| setpoint (A) and less than or equal to the high setpoint (B).(2) |
|
|
|
Out of Range | 3 | Discrete event will occur if the assigned variable is less than or equal to the low |
|
| setpoint (A) or greater than or equal to the high setpoint (B).(2) |
|
|
|
(1)An event does not occur if the assigned variable is equal to the setpoint.
(2)An event occurs if the assigned variable is equal to the setpoint.
Table 8-6 Event actions
|
| DeviceNet |
|
ProLink II label | Display label | code | Description |
|
|
|
|
Start sensor zero | START ZERO | 1 | Initiates a zero calibration procedure |
|
|
|
|
Reset mass total | RESET MASS | 2 | Resets the value of the mass totalizer to 0 |
|
|
|
|
Reset volume total | RESET VOL | 3 | Resets the value of the liquid volume totalizer to 0(1) |
|
|
|
|
Reset gas std volume total | RESET GSV | 21 | Resets the value of the gas standard volume totalizer |
|
|
| to 0(2) |
Reset API ref vol total | RESET TCORR | 4 | Resets the value of the API |
|
|
| volume totalizer to 0(3) |
|
|
|
|
Reset ED ref vol total | RESET STD V | 5 | Resets the value of the ED standard volume totalizer |
|
|
| to 0(4) |
Reset ED net mass total | RESET NET M | 6 | Resets the value of the ED net mass totalizer to 0(4) |
|
|
|
|
Reset ED net vol total | RESET NET V | 7 | Resets the value of the ED net volume totalizer to 0(4) |
|
|
|
|
62 | Micro Motion® Model 2400S Transmitters for DeviceNet™ |