Using the Transmitter
To reset an individual totalizer or inventory:
1.Click ProLink > Totalizer Control or ProLink > ED Totalizer Control (if the enhanced density application is enabled).
2.Click the appropriate button (e.g., Reset Mass Total, Reset Volume Inventory, Reset Net Mass Total).
With a DeviceNet tool
Using a DeviceNet tool, three methods are available for totalizer and inventory control:
•EDS – If you have imported the EDS into your DeviceNet tool, you can perform the following functions from the EDS user interface:
-Reset mass totalizer
-Reset mass inventory
-Reset liquid volume totalizer
-Reset liquid volume inventory
-Reset API reference volume total
-Reset API reference volume inventory
-Reset gas standard volume totalizer
-Reset gas standard volume inventory
-Reset ED standard volume total
-Reset ED net mass total
-Reset ED net volume total
-Reset ED standard volume inventory
-Reset ED net mass inventory
-Reset ED net volume inventory
•Explicit write – Using a Set, a Reset Total, or a Reset Inventory service, you can perform the functions listed in Table
•Output assemblies – Five output assemblies are provided, supporting the functions listed in Table
Table
To accomplish this | Use this device profile data |
|
|
Stop all totalizers and inventories | Analog Input Point Object (0x0A) |
| Instance: 0 |
| Attribute ID: 100 |
| Service: Set |
| Value: 0 |
|
|
Start all totalizers and inventories | Analog Input Point Object (0x0A) |
| Instance: 0 |
| Attribute ID: 100 |
| Service: Set |
| Value: 1 |
|
|
Reset all totalizers | Analog Input Point Object (0x0A) |
| Instance: 0 |
| Attribute ID: 101 |
| Service: Set |
| Value: 1 |
|
|
Using a DeviceNet Tool
Required Configuration
Using the Transmitter
Optional Configuration
Configuration and Use Manual | 51 |