Optional Configuration
Table
|
| Default |
| Affected by |
Alarm code | ProLink II message | severity | Configurable? | fault timeout? |
|
|
|
|
|
A116 | API: Temperature Outside Standard Range | Info | Yes | No |
|
|
|
|
|
A117 | API: Density Outside Standard Range | Info | Yes | No |
|
|
|
|
|
A120 | ED: Unable to Fit Curve Data | Info | No | No |
|
|
|
|
|
A121 | ED: Extrapolation Alarm | Info | Yes | No |
|
|
|
|
|
A131 | Meter Verification/Outputs at Last Value | Info | Yes | No |
|
|
|
|
|
A132 | Simulation Mode Active | Info | Yes | No |
|
|
|
|
|
A133 | PIC UI EEPROM Error | Info | Yes | No |
|
|
|
|
|
(1) Can be set to either Informational or Ignore, but cannot be set to Fault.
8.9Configuring the display
If your transmitter has a display, you can configure a variety of parameters that control the display functionality.
8.9.1Update period
The Update Period (or Display Rate) parameter controls how often the display is refreshed with current data. The default is 200 milliseconds; the range is 100 milliseconds to 10,000 milliseconds (10 seconds).
To configure Update Period:
•Using ProLink II, see Figure
•Using the display menus, see Figure
•Using a DeviceNet tool, see Table
8.9.2Language
The display can be configured to use any of the following languages for data and menus:
•English
•French
•German
•Spanish
To set the display language:
•Using ProLink II, see Figure
•Using the display menus, see Figure
•Using a DeviceNet tool, see Table
8.9.3Enabling and disabling display functions
Table
Using a DeviceNet Tool
Required Configuration
Using the Transmitter
Optional Configuration