be configured to match pressure unit

Pressure Compensation and Temperature Compensation

Figure 9-2Configuring pressure compensation with a DeviceNet tool

 

 

 

 

 

 

Class: Calibration Object (0x65)

 

 

 

 

 

 

Instance: 1

 

 

Set pressure unit

 

 

 

 

 

 

Attribute ID: 29

 

 

 

 

 

 

 

 

 

 

Value: See Table 6-7

(1) Pressure measurement unit must

 

 

 

 

Service: Set

be configured to match pressure unit

 

 

 

 

 

used by external device. See

 

 

 

 

 

 

 

 

 

Class: Calibration Object (0x65)

Section 6.3.

 

 

 

 

 

 

 

 

 

 

Instance: 1

(2) See Section 9.4.

 

Set flow calibration

 

 

 

 

 

Attribute ID: 32

 

 

pressure

 

 

 

 

 

 

Data type: REAL

 

 

 

 

 

 

 

 

 

 

 

 

Service: Set

 

 

 

 

 

 

 

 

 

 

 

 

 

Class: Calibration Object (0x65)

 

 

 

 

 

 

Instance: 1

 

 

Set flow factor

 

 

 

 

Attribute ID: 30

 

 

 

 

 

 

 

 

 

Data type: REAL

 

 

 

 

 

 

Service: Set

 

 

 

 

 

 

 

 

 

 

 

 

 

Class: Calibration Object (0x65)

 

 

 

 

 

 

Instance: 1

 

 

Set density factor

 

 

 

 

 

 

Attribute ID: 31

 

 

 

 

 

 

 

 

 

 

Data type: REAL

 

 

 

 

 

 

Service: Set

 

 

 

 

 

 

 

 

 

 

 

 

 

Class: Calibration Object (0x65)

 

 

 

 

 

 

Instance: 1

 

 

 

 

 

 

Attribute ID: 27

 

 

Enable pressure

 

 

 

 

 

 

Data type: BOOL

 

 

compensation

 

 

 

 

 

 

Value:

 

 

 

 

 

 

· 0 = disabled

 

 

 

 

 

 

· 1 = enabled

 

 

 

 

 

 

 

 

 

 

 

 

Use static

 

 

 

 

 

 

pressure value?

 

 

 

 

 

 

 

 

No

 

 

 

 

Yes

 

 

 

 

 

 

 

 

 

 

 

 

 

Class: Calibration Object (0x65)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Instance: 1

Set up output

 

Set static value

 

 

 

 

 

Attribute ID: 28

assembly(2)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Data type: REAL

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Service: Set

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Compensation

Measurement Performance

Troubleshooting

9.3External temperature compensation

External temperature compensation can be used with the petroleum measurement application or the enhanced density application.

Note: The external temperature value is used only for calculation of the derived variable in enhanced density applications or the CTL value in petroleum measurement applications. The temperature value from the sensor is used for all other calculations that require a temperature value.

Defaults

Configuration and Use Manual

85