Emerson Process Management 2400S 11.14Checking slug flow, Compensation Measurement Performance

Models: 2400S

1 188
Download 188 pages 39.36 Kb
Page 129
Image 129
11.14Checking slug flow

Troubleshooting

Table 11-3Process variables problems and remedies continued

Symptom

Cause

Suggested remedy

 

 

 

Inaccurate density reading

Problem with process fluid

• Use standard procedures to check

 

 

quality of process fluid.

 

 

 

 

Bad density calibration factors

• Verify characterization. See

 

 

Section 6.2.

 

 

 

 

Wiring problem

• Check the sensor circuitry. See

 

 

Section 11.20.

 

 

 

 

Bad flowmeter grounding

• See Section 11.7.2.

 

 

 

 

Slug flow

• See Section 11.14.

 

 

 

 

Sensor cross-talk

• Check environment for sensor with

 

 

similar (±0.5 Hz) tube frequency.

 

 

 

 

Plugged flow tube

• Check drive gain and tube frequency.

 

 

Purge the flow tubes.

 

 

 

 

Incorrect sensor orientation

• Sensor orientation must be

 

 

appropriate to process fluid. See the

 

 

installation manual for your sensor.

 

 

 

 

RTD failure

• Check for alarm conditions and follow

 

 

troubleshooting procedure for

 

 

indicated alarm.

 

 

 

 

Physical characteristics of sensor have

• Check for corrosion, erosion, or tube

 

changed

damage. See Section 11.15.

 

 

 

Temperature reading significantly

RTD failure

• Check for alarm conditions and follow

different from process temperature

 

troubleshooting procedure for

 

 

indicated alarm.

 

 

• Verify “Use external temperature”

 

 

configuration and disable if

 

 

appropriate. See Section 9.3.

 

 

 

Temperature reading slightly different

Sensor leaking heat

• Insulate the sensor.

from process temperature

 

 

 

 

 

Unusually high density reading

Plugged flow tube

• Check drive gain and tube frequency.

 

 

Purge the flow tubes.

 

 

 

 

Incorrect K2 value

• Verify characterization. See

 

 

Section 6.2.

 

 

 

Unusually low density reading

Slug flow

• See Section 11.14.

 

 

 

 

Incorrect K2 value

• Verify characterization. See

 

 

Section 6.2.

 

 

 

Unusually high tube frequency

Sensor erosion

• Contact Micro Motion. See

 

 

Section 11.3.

 

 

 

Unusually low tube frequency

Plugged flow tube, corrosion, or erosion

• Purge the flow tubes.

 

 

• Perform meter verification. See

 

 

Section 11.15.

 

 

 

Unusually low pickoff voltages

Several possible causes

• See Section 11.19.4.

 

 

 

Unusually high drive gain

Several possible causes

• See Section 11.19.3.

 

 

 

11.14Checking slug flow

A slug flow alarm is posted whenever the measured process density is outside the configured slug flow limits (i.e., density is higher or lower than the configured normal range). Slug flow is typically caused by gas in a liquid process or liquid in a gas process. See Section 8.7 for a discussion of slug flow functionality.

Compensation

Measurement Performance

Troubleshooting

Defaults

Configuration and Use Manual

121

Page 129
Image 129
Emerson Process Management 2400S 11.14Checking slug flow, Compensation Measurement Performance, Troubleshooting, Defaults