Appendix A
Default Values and Ranges
A.1 Overview
This appendix provides information on the default values for most transmitter parameters. Where appropriate, valid ranges are also defined.
These default values represent the transmitter configuration after a master reset. Depending on how the transmitter was ordered, certain values may have been configured at the factory.
A.2 Most frequently used defaults and ranges
The table below contains the default values and ranges for the most frequently used transmitter settings.
Table A-1 Transmitter default values and ranges
Type | Setting | Default | Range | Comments |
|
|
|
|
|
Flow | Flow direction | Forward |
|
|
|
|
|
|
|
| Flow damping | 0.64 sec | ||
|
|
|
| corrected to nearest lower |
|
|
|
| value in list of preset values. |
|
|
|
| For gas applications, |
|
|
|
| Micro Motion recommends a |
|
|
|
| minimum value of 2.56. |
|
|
|
|
|
| Flow calibration factor | 1.00005.13 |
| For |
|
|
|
| value represents the FCF and |
|
|
|
| FT factors concatenated. See |
|
|
|
| Section 6.2.2. |
|
|
|
|
|
| Mass flow units | g/s |
|
|
|
|
|
|
|
| Mass flow cutoff | 0.0 g/s |
| Recommended setting is 5% |
|
|
|
| of the sensor’s rated |
|
|
|
| maximum flowrate. |
|
|
|
|
|
| Volume flow type | Liquid volume |
|
|
|
|
|
|
|
| Volume flow units | L/s |
|
|
|
|
|
|
|
| Volume flow cutoff | 0/0 L/s | x is obtained by multiplying | |
|
|
|
| the flow calibration factor by |
|
|
|
| 0.2, using units of liters per |
|
|
|
| second. |
|
|
|
|
|
Meter factors | Mass factor | 1.00000 |
|
|
|
|
|
|
|
| Density factor | 1.00000 |
|
|
|
|
|
|
|
| Volume factor | 1.00000 |
|
|
|
|
|
|
|
Compensation
Measurement Performance
Troubleshooting
Defaults
Configuration and Use Manual | 131 |