Emerson Process Management 2400S manual Erratic non-zeroflow rate under

Models: 2400S

1 188
Download 188 pages 39.36 Kb
Page 128
Image 128
Erratic non-zero flow rate under

Troubleshooting

Table 11-3Process variables problems and remedies continued

Symptom

Cause

Suggested remedy

 

 

 

Erratic non-zero flow rate under

Leaking valve or seal

• Check pipeline.

no-flow conditions

 

 

Slug flow

• See Section 11.14.

 

 

 

 

 

Plugged flow tube

• Check drive gain and tube frequency.

 

 

Purge the flow tubes.

 

 

 

 

Incorrect sensor orientation

• Sensor orientation must be

 

 

appropriate to process fluid. See the

 

 

installation manual for your sensor.

 

 

 

 

Wiring problem

• Check the sensor circuitry. See

 

 

Section 11.20.

 

 

 

 

Vibration in pipeline at rate close to

• Check environment and remove

 

sensor tube frequency

source of vibration.

 

 

 

 

Damping value too low

• Check configuration. See Section 8.4.

 

 

 

 

Mounting stress on sensor

• Check sensor mounting. Ensure:

 

 

- Sensor is not being used to support

 

 

pipe.

 

 

- Sensor is not being used to correct

 

 

pipe misalignment.

 

 

- Sensor is not too heavy for pipe.

 

 

 

 

Sensor cross-talk

• Check environment for sensor with

 

 

similar (±0.5 Hz) tube frequency.

Erratic non-zero flow rate when flow is steady

Slug flow

• See Section 11.14.

 

 

Damping value too low

• Check configuration. See Section 8.4.

 

 

Plugged flow tube

• Check drive gain and tube frequency.

 

Purge the flow tubes.

 

 

Excessive or erratic drive gain

• See Section 11.19.3.

 

 

Output wiring problem

• Verify wiring between transmitter and

 

receiving device. See the installation

 

manual for your transmitter.

 

Problem with receiving device

• Test with another receiving device.

 

Wiring problem

• Check the sensor circuitry. See

 

 

Section 11.20.

 

 

 

Inaccurate flow rate or batch total

Bad flow calibration factor

• Verify characterization. See

 

 

Section 6.2.

 

 

 

 

Inappropriate measurement unit

• Check configuration. See

 

 

Section 11.16.

 

 

 

 

Bad sensor zero

• Rezero the flowmeter or restore the

 

 

factory zero or prior zero. See

 

 

Section 10.5.

 

 

 

 

Bad density calibration factors

• Verify characterization. See

 

 

Section 6.2.

 

 

 

 

Bad flowmeter grounding

• See Section 11.7.2.

 

 

 

 

Slug flow

• See Section 11.14.

 

 

 

 

Wiring problem

• Check the sensor circuitry. See

 

 

Section 11.20.

 

 

 

120

Micro Motion® Model 2400S Transmitters for DeviceNet

Page 128
Image 128
Emerson Process Management 2400S manual Erratic non-zeroflow rate under