Measurement Performance
Figure
Close shutoff valve
downstream from sensor
D3 calibration
Fill sensor with
D3 fluid
D4 calibration
Fill sensor with
D4 fluid
Enter density of
D3 fluid
Start D3
calibration
Monitor status
Object: Calibration Object (0x65)
Instance: 1
Attribute ID: 15
Data type: REAL
Service: Set
Object: Calibration Object (0x65)
Instance: 1
Service: 0x4F
Object: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 16
Bit: 7
Service: Get
Enter density of
D4 fluid
Start D4
calibration
Monitor status
Object: Calibration Object (0x65)
Instance: 1
Attribute ID: 16
Data type: REAL
Service: Set
Object: Calibration Object (0x65)
Instance: 1
Service: 0x50
Object: Diagnostics Object (0x66)
Instance: 1
Attribute ID: 16
Bit: 6
Service: Get
No |
| Bit 7 0ff? |
| Yes |
|
|
|
|
No | Bit 6 0ff? |
Yes
Done | Done |
108 | Micro Motion® Model 2400S Transmitters for DeviceNet™ |