With ProLink II

Using the Transmitter

With ProLink II

To view current totals for the totalizers and inventories with ProLink II:

1.Click ProLink.

2.Select Process Variables, API Process Variables, or ED Process Variables.

With a DeviceNet tool

To view current totals for the totalizers and inventories with a DeviceNet tool, see Section 7.3.3.

7.7.2Controlling totalizers and inventories

Specific starting, stopping, and resetting functionality depends on the tool you are using.

With the display

If the required value is shown on the display, you can use the display to start and stop all totalizers and inventories simultaneously, or to reset individual totalizers. See the flowchart in Figure 7-3.You cannot reset any inventories with the display.

Figure 7-3Controlling totalizers and inventories with the display

Process variable

display

Scroll

API total(1)(2)

Mass total(1)

Scroll

Volume total(1)

Scroll

ED total(1)(2)

Select

 

 

 

 

 

 

 

 

 

 

E1--SP(3)

 

 

 

 

 

 

EXIT

Scroll

Scroll

STOP/START(4)(5)

Scroll

RESET(6)(7)

Scroll

 

 

E2--SP(3)

 

 

 

 

 

 

 

 

 

Select

 

 

Select

 

 

 

 

 

STOP/START YES?

 

RESET YES?

 

 

 

Yes

No

 

Yes

No

 

 

 

Select

Scroll

 

Select

 

Scroll

(1)Displayed only if configured as a display variable.

(2)The petroleum measurement application or enhanced density application must be enabled.

(3)The Event Setpoint screens can be used to define or change Setpoint A for Event 1 or Event 2 only. These screens are displayed only for specific types of events. To change the setpoint for an event defined on mass total, you must enter the totalizer management menu from the mass total screen. To change the setpoint for an event defined on volume total, you must enter the totalizer management menu from the volume total screen. See Section 8.6.3 for more information.

(4)The display must be configured to allow stopping and starting. See Section 8.9.3.

(5)All totalizers and inventories will be stopped and started together, including API and enhanced density totalizers and inventories.

(6)The display must be configured to allow totalizer resetting. See Section 8.9.3.

(7)Only the totalizer currently shown on the display will be reset. No other totalizers will be reset, and no inventories will be reset. Be sure that the totalizer you want to reset is displayed before performing this reset.

Using a DeviceNet Tool

Required Configuration

Using the Transmitter

Optional Configuration

Configuration and Use Manual

49

Page 57
Image 57
Emerson Process Management 2400S manual With ProLink, With a DeviceNet tool, 7.7.2Controlling totalizers and inventories