Troubleshooting

11.19.1Obtaining the test point values

To obtain the test point values:

With the display, configure the required test points as display variables. See Section 8.9.5.

With ProLink II

a.Click ProLink > Diagnostic Information.

b.Observe or record the values displayed for Tube Frequency, Left Pickoff, Right Pickoff, and Drive Gain.

With a DeviceNet tool, execute a Get for the attributes listed in Table 11-4.

Table 11-4Test points with DeviceNet tool

Test point

Class

Instance

Attribute

 

 

 

 

Drive gain

Diagnostics Object (0x66)

1

20

 

 

 

 

Tube period

 

 

21

 

 

 

 

Left pickoff

 

 

23

 

 

 

 

Right pickoff

 

 

24

 

 

 

 

11.19.2Evaluating the test points

Use the following guidelines to evaluate the test points:

If the drive gain is erratic, negative, or saturated, refer to Section 11.19.3.

If the value for the left or right pickoff does not equal the appropriate value from Table 11-5,based on the sensor flow tube frequency, refer to Section 11.19.4.

If the values for the left and right pickoffs equal the appropriate values from Table 11-5,based on the sensor flow tube frequency, record your troubleshooting data and contact Micro Motion customer service. See Section 11.3.

Table 11-5Sensor pickoff values

Sensor(1)

Pickoff value

ELITE® CMF sensors

3.4 mV peak-to-peak per Hz based on sensor flow tube frequency

 

 

F025, F050, F100 sensors

3.4 mV peak-to-peak per Hz based on sensor flow tube frequency

 

 

F200 sensors

2.0 mV peak-to-peak per Hz based on sensor flow tube frequency

 

 

H025, H050, H100 sensors

3.4 mV peak-to-peak per Hz based on sensor flow tube frequency

 

 

H200 sensors

2.0 mV peak-to-peak per Hz based on sensor flow tube frequency

 

 

R025, R050, or R100 sensors

3.4 mV peak-to-peak per Hz based on sensor flow tube frequency

 

 

R200 sensors

2.0 mV peak-to-peak per Hz based on sensor flow tube frequency

 

 

T-Series sensors

0.5 mV peak-to-peak per Hz based on sensor flow tube frequency

 

 

CMF400 I.S. sensors

2.7 mV peak-to-peak per Hz based on sensor flow tube frequency

 

 

(1) If your sensor is not listed, contact Micro Motion. See Section 11.3.