7.3.3With a DeviceNet tool

Using the Transmitter

To view enhanced density process variables (if the enhanced density application is enabled), click ProLink > ED Process Variables. Different enhanced density process variables are displayed, depending on the configuration of the enhanced density application.

7.3.3With a DeviceNet tool

There are two methods that can be used to view process variables with a DeviceNet tool:

You can execute Gets to read the current values of individual process variables from the appropriate objects. Table 7-1lists the most commonly used process variables, by class, instance, attribute, and data type. For more information, see the manual entitled Micro Motion Model 2400S Transmitters for DeviceNet: Device Profile.

You can use the predefined input assemblies. The predefined input assemblies are summarized in Table 7-2.For more information, see the manual entitled Micro Motion Model 2400S Transmitters for DeviceNet: Device Profile.

Table 7-1Process data in DeviceNet objects

 

 

Attribute

Data

 

Class

Instance

ID

type

Description

 

 

 

 

 

Analog Input Point

1 (mass)

3

REAL

Mass flow rate

Object (0x04)

 

 

 

 

 

100

REAL

Mass total

 

 

 

 

 

 

 

 

 

101

REAL

Mass inventory

 

 

 

 

 

 

 

102

UINT

Mass flow measurement unit

 

 

 

 

 

 

 

103

UINT

Mass total and mass inventory

 

 

 

 

measurement unit

 

 

 

 

 

 

2 (liquid volume)

3

REAL

Liquid volume flow rate

 

 

 

 

 

 

 

100

REAL

Liquid volume total

 

 

 

 

 

 

 

101

REAL

Liquid volume inventory

 

 

 

 

 

 

 

102

UINT

Liquid volume flow measurement unit

 

 

 

 

 

 

 

103

UINT

Liquid volume total and liquid volume inventory

 

 

 

 

measurement unit

 

 

 

 

 

 

3 (density)

3

REAL

Density

 

 

 

 

 

 

 

102

UINT

Density measurement unit

 

 

 

 

 

 

4 (temperature)

3

REAL

Temperature

 

 

 

 

 

 

 

102

UINT

Temperature measurement unit

 

 

 

 

 

Gas Standard Volume

1 (gas standard

1

REAL

Gas standard volume flow rate

Object (0x64)

volume)

 

 

 

2

REAL

Gas standard volume total

 

 

 

 

 

 

 

 

 

3

REAL

Gas standard volume inventory

 

 

 

 

 

 

 

5

REAL

Gas standard volume flow measurement unit

 

 

 

 

 

 

 

6

REAL

Gas standard volume total and gas standard

 

 

 

 

volume inventory measurement unit

 

 

 

 

 

Using a DeviceNet Tool

Required Configuration

Using the Transmitter

Optional Configuration

Configuration and Use Manual

37