Using the Transmitter
To view enhanced density process variables (if the enhanced density application is enabled), click ProLink > ED Process Variables. Different enhanced density process variables are displayed, depending on the configuration of the enhanced density application.
7.3.3With a DeviceNet tool
There are two methods that can be used to view process variables with a DeviceNet tool:
•You can execute Gets to read the current values of individual process variables from the appropriate objects. Table
•You can use the predefined input assemblies. The predefined input assemblies are summarized in Table
Table 7-1 Process data in DeviceNet objects
|
| Attribute | Data |
| |
Class | Instance | ID | type | Description | |
|
|
|
|
| |
Analog Input Point | 1 (mass) | 3 | REAL | Mass flow rate | |
Object (0x04) |
|
|
|
| |
| 100 | REAL | Mass total | ||
|
| ||||
|
|
|
|
| |
|
| 101 | REAL | Mass inventory | |
|
|
|
|
| |
|
| 102 | UINT | Mass flow measurement unit | |
|
|
|
|
| |
|
| 103 | UINT | Mass total and mass inventory | |
|
|
|
| measurement unit | |
|
|
|
|
| |
| 2 (liquid volume) | 3 | REAL | Liquid volume flow rate | |
|
|
|
|
| |
|
| 100 | REAL | Liquid volume total | |
|
|
|
|
| |
|
| 101 | REAL | Liquid volume inventory | |
|
|
|
|
| |
|
| 102 | UINT | Liquid volume flow measurement unit | |
|
|
|
|
| |
|
| 103 | UINT | Liquid volume total and liquid volume inventory | |
|
|
|
| measurement unit | |
|
|
|
|
| |
| 3 (density) | 3 | REAL | Density | |
|
|
|
|
| |
|
| 102 | UINT | Density measurement unit | |
|
|
|
|
| |
| 4 (temperature) | 3 | REAL | Temperature | |
|
|
|
|
| |
|
| 102 | UINT | Temperature measurement unit | |
|
|
|
|
| |
Gas Standard Volume | 1 (gas standard | 1 | REAL | Gas standard volume flow rate | |
Object (0x64) | volume) |
|
|
| |
2 | REAL | Gas standard volume total | |||
|
| ||||
|
|
|
|
| |
|
| 3 | REAL | Gas standard volume inventory | |
|
|
|
|
| |
|
| 5 | REAL | Gas standard volume flow measurement unit | |
|
|
|
|
| |
|
| 6 | REAL | Gas standard volume total and gas standard | |
|
|
|
| volume inventory measurement unit | |
|
|
|
|
|
Using a DeviceNet Tool
Required Configuration
Using the Transmitter
Optional Configuration
Configuration and Use Manual | 37 |