Optional Configuration
Table 8-6 Event actions continued
|
| DeviceNet |
|
ProLink II label | Display label | code | Description |
|
|
|
|
Reset all totals | RESET ALL | 8 | Resets the value of all totalizers to 0 |
|
|
|
|
Start/stop all totalization | START STOP | 9 | If totalizers are running, stops all totalizers |
|
|
| If totalizers are not running, starts all totalizers |
|
|
|
|
Increment current ED | INCR CURVE | 18 | Changes the active enhanced density curve from 1 to |
curve |
|
| 2, from 2 to 3, etc.(4) |
|
|
|
|
(1)Displayed only if Volume Flow Type = Liquid.
(2)Displayed only if Volume Flow Type = Gas.
(3)Available only if the petroleum measurement application is installed.
(4)Available only if the enhanced density application is installed.
Example | Define Event 1 to be active when the mass flow rate in forward or |
| backward direction is less than 2 lb/min. Additionally, if this occurs, all |
| totalizers should be stopped. |
Using ProLink II:
1.Specify lb/min as the mass flow unit. See Section 6.3.1.
2.Configure the Flow Direction parameter for bidirectional flow. See Section 8.5.
3.Select Event 1.
4.Configure:
•Event Type = Low
•Process Variable (PV) = Mass Flow Rate
•Low Setpoint (A) = 2
5.In the Discrete Input panel, open the dropdown list for Start/Stop All Totalization and select Discrete Event 1.
Using a DeviceNet tool:
1.Specify lb/min as the mass flow unit. See Section 6.3.1.
2.Configure the Flow Direction parameter for bidirectional flow. See Section 8.5.
3.In the Diagnostics Object (0x66), Instance 1, set the following attributes:
•Discrete event index (Attribute 6) = 0
•Discrete event action (Attribute 7) = 1
•Discrete event process variable (Attribute 10) = 0
•Discrete event setpoint A (Attribute 8) = 2
•Discrete event action code (Attribute 84) = 9
•Discrete event assignment (Attribute 85) = 57
Using a DeviceNet Tool
Required Configuration
Using the Transmitter
Optional Configuration