Connecting with ProLink II or Pocket ProLink Software

WARNING

Removing the transmitter housing cover in a hazardous area can cause an explosion.

Because the transmitter housing cover must be removed to connect to the service port clips, the service port clips should be used only for temporary connections, e.g., for configuration or troubleshooting purposes.

When the transmitter is in an explosive atmosphere, use a different method to connect to your transmitter.

Figure 4-1Serial port connections to service port clips

PC

Service port clips

RS-485/A RS-485/B

25-pin to 9-pin serial port adapter (if necessary)

RS-485 to RS-232 signal converter

3.Start ProLink II or Pocket ProLink. In the Connection menu, click Connect to Device. In the screen that appears, specify:

Protocol: Service Port

COM Port: as appropriate

No other parameters are required.

4.Click Connect. The software will attempt to make the connection.

5.If an error message appears:

a.Swap the leads between the two service port clips and try again.

b.Ensure that you are using the correct COM port.

c.Check all the wiring between the PC and the transmitter.

d.Verify the RS-485 to RS-232 signal converter.

Before You Begin