Emerson Process Management 2400S manual 10.7Performing temperature calibration

Models: 2400S

1 188
Download 188 pages 39.36 Kb
Page 117
Image 117
10.7Performing temperature calibration

Measurement Performance

10.7Performing temperature calibration

Temperature calibration is a two-part procedure: temperature offset calibration and temperature slope calibration. The entire procedure must be completed without interruption.

To perform temperature calibration, you must use ProLink II. See Figure 10-12.

Compensation

Figure 10-12Temperature calibration – ProLink II

Temperature Offset calibration

Fill sensor with low-

temperature fluid

Wait until sensor achieves

thermal equilibrium

ProLink Menu >

Calibration >

Temp offset cal

Enter temperature of low-

temperature fluid

Do Cal

Calibration in Progress

light turns red

Calibration in Progress

light turns green

Close

Temperature Slope calibration

Fill sensor with high-

temperature fluid

Wait until sensor achieves

thermal equilibrium

ProLink Menu >

Calibration >

Temp slope cal

Enter temperature of high-

temperature fluid

Do Cal

Calibration in Progress

light turns red

Calibration in Progress

light turns green

Close

Done

Measurement Performance

Troubleshooting

Defaults

Configuration and Use Manual

109

Page 117
Image 117
Emerson Process Management 2400S manual 10.7Performing temperature calibration, 12 Temperature calibration – ProLink