Measurement Performance
10.7Performing temperature calibration
Temperature calibration is a
To perform temperature calibration, you must use ProLink II. See Figure
Compensation
Figure 10-12 Temperature calibration – ProLink II
Temperature Offset calibration |
Fill sensor with low- |
temperature fluid |
Wait until sensor achieves |
thermal equilibrium |
ProLink Menu > |
Calibration > |
Temp offset cal |
Enter temperature of low- |
temperature fluid |
Do Cal |
Calibration in Progress |
light turns red |
Calibration in Progress |
light turns green |
Close |
Temperature Slope calibration |
Fill sensor with high- |
temperature fluid |
Wait until sensor achieves |
thermal equilibrium |
ProLink Menu > |
Calibration > |
Temp slope cal |
Enter temperature of high- |
temperature fluid |
Do Cal |
Calibration in Progress |
light turns red |
Calibration in Progress |
light turns green |
Close |
Done |
Measurement Performance
Troubleshooting
Defaults
Configuration and Use Manual | 109 |