Pressure Compensation and Temperature Compensation

Figure 9-4Configuring external temperature compensation with a DeviceNet tool

(1) Temperature measurement unit

 

 

 

 

Class: Analog Input Point Object (0x0A)

 

 

 

 

Instance: 4

Set temperature unit

 

 

must be configured to match

 

 

Attribute ID: 102

 

 

temperature unit used by

 

 

 

 

Value: See Table 6-6

external device. See

 

 

 

 

Service: Set

Section 6.3.

 

 

 

 

 

(2) See Section 9.4.

 

 

 

 

 

 

 

 

 

Class: Calibration Object (0x65)

 

 

 

 

 

 

 

 

 

 

Instance: 1

 

 

 

 

 

Attribute ID: 25

 

Enable temperature

 

 

 

Data type: BOOL

 

compensation

 

 

 

 

 

Value:

 

 

 

 

 

· 0 = disabled

 

 

 

 

 

· 1 = enabled

 

 

 

 

 

 

 

 

 

 

 

Use static

 

 

 

 

 

 

temp value?

 

 

 

 

 

 

 

 

No

 

 

 

 

Yes

 

 

 

 

 

 

 

 

 

 

 

 

 

Class: Calibration Object (0x65)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Instance: 1

Set up output

 

Set static value

 

 

 

 

 

Attribute ID: 26

assembly(2)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Data type: REAL

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Service: Set

 

 

 

 

 

 

 

 

 

 

 

 

 

 

9.4Obtaining external pressure and temperature data

The DeviceNet output assemblies used to obtain external pressure and/or temperature data are listed in Table 9-1.Use standard DeviceNet methods to implement the required connection.

Table 9-1 Output assemblies used for pressure or temperature compensation

Instance ID

Data description

Size

Bytes

Data type

 

 

 

 

 

50

External pressure

4 bytes

Bytes 0–3

REAL

 

 

 

 

 

51

External temperature

4 bytes

Bytes 0–3

REAL

 

 

 

 

 

52

External pressure

8 bytes

Bytes 0–3

REAL

 

External temperature

 

Bytes 4–7

REAL

 

 

 

 

 

Compensation

Measurement Performance