Pressure Compensation and Temperature Compensation
Figure
(1) Temperature measurement unit |
|
|
|
| Class: Analog Input Point Object (0x0A) |
|
|
|
| Instance: 4 | |
Set temperature unit |
|
| |||
must be configured to match |
|
| Attribute ID: 102 | ||
|
| ||||
temperature unit used by |
|
|
|
| Value: See Table |
external device. See |
|
|
|
| Service: Set |
Section 6.3. |
|
|
|
|
|
(2) See Section 9.4. |
|
|
|
|
|
|
|
|
| Class: Calibration Object (0x65) | |
|
|
|
|
| |
|
|
|
|
| Instance: 1 |
|
|
|
|
| Attribute ID: 25 |
| Enable temperature | ||||
|
|
| Data type: BOOL | ||
| compensation |
|
| ||
|
|
| Value: | ||
|
|
|
|
| · 0 = disabled |
|
|
|
|
| · 1 = enabled |
|
|
|
|
|
|
|
|
|
|
| Use static |
| |||||||
|
|
|
|
| temp value? |
|
|
|
|
|
|
| |
| No |
|
|
|
| Yes |
|
|
|
| |||
|
|
|
|
|
|
|
|
| Class: Calibration Object (0x65) | ||||
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
| Instance: 1 |
Set up output |
| Set static value |
|
| |||||||||
|
|
| Attribute ID: 26 | ||||||||||
assembly(2) |
|
|
| ||||||||||
|
|
| |||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
| Data type: REAL |
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
| Service: Set |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
9.4Obtaining external pressure and temperature data
The DeviceNet output assemblies used to obtain external pressure and/or temperature data are listed in Table
Table
Instance ID | Data description | Size | Bytes | Data type |
|
|
|
|
|
50 | External pressure | 4 bytes | Bytes | REAL |
|
|
|
|
|
51 | External temperature | 4 bytes | Bytes | REAL |
|
|
|
|
|
52 | External pressure | 8 bytes | Bytes | REAL |
| External temperature |
| Bytes | REAL |
|
|
|
|
|